The 

HI 95706 

is an auto diagnostic portable microprocessor meter that benefits from 

HANNA

’s

years of experience as a manufacturer of analytical instruments. It has the advanced optical system
based on a special tungsten lamp and a narrow band interference filter that allows most accurate and

repeatable readings. All instruments are factory calibrated and the electronic and optical design
minimizes the need of frequent calibration.
With the powerful CAL CHECK™ 

validation

 function you are able to validate good performance of

your instrument at any time. The validation procedure is extremely user friendly. Just use the exclusive

HANNA

 ready-made, NIST traceable standards to verify the performance of the instrument and

recalibrate if necessary.
All instruments are splash waterproof and the lamp and filter units are protected from dust or dirt
by a transparent cup. This makes the instruments fulfill field applications. Display codes aid the user

in routine operation. The meter has an auto-shut off feature that will turn off the instrument  after
10 minutes of non use or after 1 hour if left in the 

calibration mode

.

The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same position
every time it is placed into the measurement cell. It is designed to fit a cuvet with a larger neck

making it easier to add both sample and reagents. The cuvet is made from special optical glass to
obtain best results.
The 

HI 95706

 meter measures the phosphorus (P) content in water samples in the 0.0 to 15.0 mg/L

(ppm) range. The method is an adaptation from 

Standard Method for the Examination of Water and

Wastewater

.

The reagents are in

 

powder and liquid form and are supplied in packets and bottles. The amount of

reagent is precisely dosed to ensure the maximum repeatability.

degree Celsius
degree Fahrenheit
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter

°C

:

°F

:

mg/L

:

mL

:

GENERAL DESCRIPTION

ABBREVIATIONS

17

or

• Press READ/TIMER and “SIP” will blink during

measurement.

• The instrument will show for three seconds the

CAL CHECK™ standard value.
Then the date of last calibration (e.g.: “01.08”)

appears on the display, or “01.01” if the factory
calibration was selected. In both cases the

number of the month is blinking, ready for
date input.

Note

: if the display shows “ERR” the calibration

procedure failed. Verify that the CAL CHECK™

Standards HI 95706-11 Cuvets, 

A

 and

 B

, are

free from fingerprints or dirt and that they are
inserted correctly.

DATE INPUT

• Keep READ/TIMER pressed to scroll to the

desired month number (01-12).

• When the correct month has been set, press

ZERO to confirm. Now the display will show the
day number blinking. Keep READ/TIMER pressed
to scroll to the desired day number (01-31).

Note

: it is possible to change from 

day 

to

month 

and vice versa by pressing ZERO.

• When both the day and month have been

selected, hold CAL CHECK 

for three seconds

 to

store date and calibration values. The instrument
will show for one second “Stor”, to confirm

that the new calibration data has been accepted.

• The instrument will return automatically to the

measurement mode by  displaying "706" on
the LCD.

Summary of Contents for HI 95706C

Page 1: ...1 Instruction Manual HI 95706C Phosphorus ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95706R2 06 06 w w w h a n n a i n s t c o m...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Page 3: ...ACCESSORIES HI 95706 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuv...

Page 4: ...asures the phosphorus P content in water samples in the 0 0 to 15 0 mg L ppm range The method is an adaptation from Standard Method for the Examination of Water and Wastewater The reagents are in powd...

Page 5: ...tandardsolutionof6 00 mg L phosphorus and a representative lot of reagents an operator obtained with a single instrumentastandarddeviationof0 2mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...ever possible use the same cuvet for both It is necessary that the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of l...

Page 8: ...ation mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCED...

Page 9: ...ration is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a...

Page 10: ...ure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zer...

Reviews: