Hanna Instruments HI 95706C Instruction Manual Download Page 10

10

The flashing digits indicate that:

1) the month can be set
2) the day can be set

“Storage”: this prompt appears for 1 second at the end of the

calibration procedure

, to indicate that the calibration data has

been stored.

“Error”: the concentration of the calibration solution used is not

correct. Repeat the 

calibration procedure

 with the right standard

solution, and verify it is not expired. If the calibration procedure fails
again, contact your dealer or the nearest Hanna Customer Service

Center.

ERROR  MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-

to-noise ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check

that the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement.
Please check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check
if the right sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in
the 

measurement procedure

 for zeroing the meter.

11

Under range. A blinking "0.00" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make

sure you use the same cuvet for reference (zero) and measurement.

A  flashing value of the maximum concentration indicates an

over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.

A flashing value lower than the maximum concentration indicates
a  low signal-to-noise ratio condition. In this case accuracy of the

result is not guaranteed. Repeat the 

measurement procedure

.

The instructions listed below should be carefully followed during testing to ensure best accuracy.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

• Suspended matter in large amounts may cause

interference, therefore this should be removed by

prior filtration.

• For a correct filling of the cuvet: the liquid in the

cuvet forms a convexity on the top; the bottom of
this convexity must be at the same level of the 10

mL mark.

• It is important that the sample does not contain any

debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of

fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 (tissue for wiping cuvets, see chapter

ACCESSORIES) or a lint-free cloth prior to insertion.

Summary of Contents for HI 95706C

Page 1: ...1 Instruction Manual HI 95706C Phosphorus ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95706R2 06 06 w w w h a n n a i n s t c o m...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Page 3: ...ACCESSORIES HI 95706 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuv...

Page 4: ...asures the phosphorus P content in water samples in the 0 0 to 15 0 mg L ppm range The method is an adaptation from Standard Method for the Examination of Water and Wastewater The reagents are in powd...

Page 5: ...tandardsolutionof6 00 mg L phosphorus and a representative lot of reagents an operator obtained with a single instrumentastandarddeviationof0 2mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...ever possible use the same cuvet for both It is necessary that the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of l...

Page 8: ...ation mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCED...

Page 9: ...ration is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a...

Page 10: ...ure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zer...

Reviews: