Instruction Manual

HI 95706C

Phosphorus ISM

w w w . h a n n a i n s t . c o m

This Instrument is in

Compliance with the CE Directives

20

MAN95706R2   06/06

w w w . h a n n a i n s t . c o m

Summary of Contents for HI 95706C

Page 1: ...1 Instruction Manual HI 95706C Phosphorus ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95706R2 06 06 w w w h a n n a i n s t c o m...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Page 3: ...ACCESSORIES HI 95706 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuv...

Page 4: ...asures the phosphorus P content in water samples in the 0 0 to 15 0 mg L ppm range The method is an adaptation from Standard Method for the Examination of Water and Wastewater The reagents are in powd...

Page 5: ...tandardsolutionof6 00 mg L phosphorus and a representative lot of reagents an operator obtained with a single instrumentastandarddeviationof0 2mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...ever possible use the same cuvet for both It is necessary that the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of l...

Page 8: ...ation mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCED...

Page 9: ...ration is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a...

Page 10: ...ure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zer...