Hanna Instruments HI 93551R Instruction Manual Download Page 8

15

14

The average mode will calculate the average for a period of up to
24 hours. If a different mode is selected for that channel, the
average will no longer calculate and the value will be lost.
At the end of the 24 hour period, the "AVG" tag will blink to
indicate that sampling has stopped, and the displayed value is
the last average calculated.
If a probe is removed or over-range is reached while in average
mode, the display shows flashing dashes and blinking "AVG" tag.
The average value is lost and will not restore even if the condition
restores. To start the average cycle again, either press CLR or re-
enter the average mode.

• While in T1-T2 (not for 

HI 93551R

), pressing DISP will switch

between 4 displays:

Note

: Pressing T1/T2 to change the channel will not alter the

information chosen to be displayed for each channel (DISP).
For example, if T1 is in average mode, the average will continue
to be calculated even if the user press T1/T2 to see T2.

Note

: Relative/average modes cannot be entered if dashes are shown

in the main portion of the LCD.

Note

: The auto-off time is disabled whenever a channel is set to

average mode.

SERIAL PORT COMMUNICATION

(HI 93551R & HI 93552R only)

HI 93551R

 and 

HI 93552R

 feature an RS232 output for transfer-

ring measurement data (once every second for 

HI 93551R

 and once

every 2 seconds for 

HI 93552R

) to devices provided with RS232 input

(PC or printer).
The communication protocol has been designed to transmit to the
receiving unit all the displayed information.
The communication is unidirectional (meter to receiving unit only) and
the transmissions consist in a 32-character ASCII string, compatible
with our optional 

HI 92000

 software.

The 32-character data string is structured as follows:

Main portion of the LCD

• byte 0

probe type: K-thermocouple (k)

• bytes 1, 2

measurement channel: T1, T2, T1-T2 (Td)

•  byte  3

measurement mode: Normal (blank), Relative (R),
Average (A), Average done (a)

• byte 4

operating mode: Hold (H), Memory recall (M)

• byte 5

blank character

• bytes 6-10

measurement: reading (XXX.X or blank XXXX),
over-range (OVRG blank), no data (blank ----)

• byte 11

temperature unit: C, F

• byte 12

blank character

Secondary LCD, left portion

• bytes 13, 14

info description: low temperature (Lo), T1 (T1)

• byte 15

blank character

• bytes 16-20

measurement: reading (XXX.X or blank XXXX),
over-range (5 blanks), no data (blank ----)

• byte 21

blank character

Secondary LCD, right portion

• bytes 22, 23

info description: high temperature (Hi), T2 (T2)

• byte 24

blank character

• bytes 25-29

measurement: reading (XXX.X or blank XXXX),
over-range (5 blanks), no data (blank ----)

• byte 30

<CR>

• byte 31

line feed

Summary of Contents for HI 93551R

Page 1: ...Instruction Manual HI 93551 HI 93551N HI 93551R HI 93542 HI 93552 HI 93552R K J T Thermometers w w w h a n n a i n s t c o m...

Page 2: ...ch as calibration of meter and probe at 0 C backlit display and capability to store and recall a reading Moreover the HI 93551R and HI 93552R models are equipped with a serial port interface WARRANTY...

Page 3: ...to 0 1 F 0 3 F 328 0 to 270 1 F Accuracy 0 5 C 100 0 to 999 9 C 1 C outside 20 C 68 F 1 F 148 0 to 999 9 F 1 5 F outside for one year excluding probe error Typical EMC Deviation 3 C 6 F Battery 3x1 5V...

Page 4: ...t the thermocouple type ALT BL toggle the backlight ON and OFF SPECIFICATIONS Range K 200 0 to 999 9 C 1000 to 1371 C 328 0 to 999 9 F 1000 to 2500 F J 200 0 to 999 9 C 328 0 to 999 9 F 1000 to 1832 F...

Page 5: ...ght ON and OFF SPECIFICATIONS Range K 200 0 to 999 9 C 1000 to 1371 C 328 0 to 999 9 F 1000 to 2500 F J 200 0 to 999 9 C 328 0 to 999 9 F 1000 to 1832 F T 200 0 to 400 0 C 328 0 to 752 0 F Resolution...

Page 6: ...this feature set the internal slide switch located in the battery compart ment see figure on page 10 OPERATIONAL GUIDE To switch the instrument ON press the ON OFF key The thermometer will carry out a...

Page 7: ...ags light up to inform the user To select the desired channel use the T1 T2 key The 2 channel and HI 93551R models are also provided with the DISP function which allows the user to select the informat...

Page 8: ...HI 93551R and HI 93552R feature an RS232 output for transfer ring measurement data once every second for HI 93551R and once every 2 seconds for HI 93552R to devices provided with RS232 input PC or pr...

Page 9: ...your nearest Hanna Customer Service Center When the battery level is below 5 a warning symbol will blink on the LCD to indicate a low battery condition If the battery level is low enough to cause err...

Page 10: ...n long probe max 250 C 482 F HI 766TV1 Pipe clamp probe max 200 C 390 F with detachable handle mini connector to be used in conjunction with the HI 766HD probe handle HI 766PA Roller surface probe max...

Page 11: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax...

Reviews: