Hanna Instruments HI 93551R Instruction Manual Download Page 7

13

12

The 

HI 93551R, HI 93552

 and 

HI 93552R

 allow the user to select the

auto-off time period through the front keyboard; press (ALT+) AOFF to
enter the mode, then set the desired auto shut-off interval (8 min, 60
min, or disabled) with the ALT key; release all buttons, wait a few
seconds and the meter will return to normal measurement mode.

BACKLIGHT FEATURE

The 

HI 93551R

HI 93552

 and 

HI 93552R

models are provided with a backlight feature,
which can be easily activated through the key-
board by pressing the (ALT+)BL keys.

Note

: The backlight automatically shuts off after approximately 1

minute with no buttons pressed.

BATTERY COMPARTMENT

See "Battery replacement" section for back cover removal/installation.

2-CHANNEL MODELS (HI 93542, HI 93552 & HI 93552R)

HI 93542

HI 93552

 and 

HI 93552R

 models can monitor two

samples through two independent temperature channels (probes).
The display shows the actual, Hi and Lo (or T1 and T2) values of the
selected channel (T1, T2 or T1-T2). The corresponding tags light up to
inform the user.
To select the desired channel, use the T1/T2 key.

The 2-channel and 

HI 93551R

 models are also provided with the

DISP function, which allows the user to select the information to be
displayed.
• While in T1 or T2, pressing DISP switches the display between

normal, relative and average measurement modes.

In the normal mode

, the main body of the LCD shows the

current temperature while the lower portion displays the Hi/Lo
limits for that channel.

When the meter enters relative mode

, the "REL" tag

lights up and the current temperature for that channel is set as
the reference temperature. Pressing CLR will also set the reference
to the current temperature.
In relative mode, the main body of the LCD shows the difference
from the reference temperature. The lower portion displays the
current temperature for the selected channel.

When average mode is entered

, the "AVG" tag lights up

and the current value is set as the new beginning value. The
average can be reset by pressing CLR while in the average mode.
In average mode, the main body of the LCD shows the average
temperature, while the lower portion displays the current tem-
perature for the selected channel.

Summary of Contents for HI 93551R

Page 1: ...Instruction Manual HI 93551 HI 93551N HI 93551R HI 93542 HI 93552 HI 93552R K J T Thermometers w w w h a n n a i n s t c o m...

Page 2: ...ch as calibration of meter and probe at 0 C backlit display and capability to store and recall a reading Moreover the HI 93551R and HI 93552R models are equipped with a serial port interface WARRANTY...

Page 3: ...to 0 1 F 0 3 F 328 0 to 270 1 F Accuracy 0 5 C 100 0 to 999 9 C 1 C outside 20 C 68 F 1 F 148 0 to 999 9 F 1 5 F outside for one year excluding probe error Typical EMC Deviation 3 C 6 F Battery 3x1 5V...

Page 4: ...t the thermocouple type ALT BL toggle the backlight ON and OFF SPECIFICATIONS Range K 200 0 to 999 9 C 1000 to 1371 C 328 0 to 999 9 F 1000 to 2500 F J 200 0 to 999 9 C 328 0 to 999 9 F 1000 to 1832 F...

Page 5: ...ght ON and OFF SPECIFICATIONS Range K 200 0 to 999 9 C 1000 to 1371 C 328 0 to 999 9 F 1000 to 2500 F J 200 0 to 999 9 C 328 0 to 999 9 F 1000 to 1832 F T 200 0 to 400 0 C 328 0 to 752 0 F Resolution...

Page 6: ...this feature set the internal slide switch located in the battery compart ment see figure on page 10 OPERATIONAL GUIDE To switch the instrument ON press the ON OFF key The thermometer will carry out a...

Page 7: ...ags light up to inform the user To select the desired channel use the T1 T2 key The 2 channel and HI 93551R models are also provided with the DISP function which allows the user to select the informat...

Page 8: ...HI 93551R and HI 93552R feature an RS232 output for transfer ring measurement data once every second for HI 93551R and once every 2 seconds for HI 93552R to devices provided with RS232 input PC or pr...

Page 9: ...your nearest Hanna Customer Service Center When the battery level is below 5 a warning symbol will blink on the LCD to indicate a low battery condition If the battery level is low enough to cause err...

Page 10: ...n long probe max 250 C 482 F HI 766TV1 Pipe clamp probe max 200 C 390 F with detachable handle mini connector to be used in conjunction with the HI 766HD probe handle HI 766PA Roller surface probe max...