SPECIFICATIONS OF

HI 93551R & HI 93552R

Keyboard Functions:

ON/OFF

 : turn the meter ON and OFF.

T1/T2

 : select the reading channel (T1, T2 or T1-T2).

HOLD

 : freeze the reading on display. The frozen value is also stored

in non-volatile memory.

CLR

 : clear the HI and LO values, reset the relative or average

measurement.

DISP

 : select Normal, Relative, Average or T1/T2 measuring mode.

ALT

 : enable the second function keys; the "ALT" tag turns on to

indicate that the second functions are enabled.

Note

: The 

ALT

 key can be released before pressing the second

function key for 1-hand operation.

ALT/ CAL

 : press and hold for about 5 seconds to enter the

Calibration mode (with reading within ±3°C range).

(ALT +) AOFF

 : set auto-off delay (8 min, 60 min, disabled).

(ALT +) °C/°F

 : change reading units (°C or °F).

(ALT +) MR

 : recall memorized value.

(ALT +) KJT

 : select the thermocouple type.

(ALT +) BL

 : toggle the backlight ON and OFF.

SPECIFICATIONS

Range (*)     K

-200.0 to 999.9°C / 1000 to 1371°C
-328.0 to 999.9°F / 1000 to 2500°F

                               

J

-200.0 to 999.9°C

-328.0 to 999.9°F / 1000 to 1832°F

                               

T

-200.0 to 400.0°C
-328.0 to 752.0°F

Resolution     K

0.1°C (-149.9 to 999.9°C)

0.2°C (-200.0 to -150.0°C) / 1°C (1000 to 1371°C)

0.1°F (-24.9 to 999.9°F) / 0.2°F (-249.9 to -25.0°F)

0.3°F (-328.0 to -250.0°F) / 1°F (1000 to 2500°F)

                               

J

0.1°C (-200.0 to 999.9°C)
0.1°F (-149.9 to 999.9°F)

0.2°F (-328.0 to -150.0°F) / 1°F (1000 to 1832°F)

                               

T

0.1°C (-149.9 to 400.0°C)

 0.2°C (-200.0 to -150.0°C)

0.1°F (0.0 to 752.0°F) / 0.2°F (-270.0 to -0.1°F)

0.3°F (-328.0 to -270.1°F)

Accuracy

±0.5°C (-100.0 to 999.9°C) / ±1°C (outside)

(@20°C/68°F)

 ±1°F (-148.0 to 999.9°F) / ±1.5°F (outside)

for one year, excluding probe error

Typical EMC Deviation

 ± 3°C / ± 6°F

Battery

3x1.5V AA (IEC LR6) batteries,

approx. 500 hours of continuous use (BL off)

Auto-off

user selectable: 8 min, 60 min, disabled

Serial Port

unidirectional, 8-bit data, 1200 baud

Communication

(see "Serial Communication" section)

Environment

-10 to 60°C (14 to 140°F); RH 100%

Dimensions

150 x 80 x 36 mm (5.9 x 3.1 x 1.4")

Weight

235 g (8.3 oz.)

(*) 

Range may be limited by probe.

Summary of Contents for HI 93551R

Page 1: ...Instruction Manual HI 93551 HI 93551N HI 93551R HI 93542 HI 93552 HI 93552R K J T Thermometers w w w h a n n a i n s t c o m...

Page 2: ...ch as calibration of meter and probe at 0 C backlit display and capability to store and recall a reading Moreover the HI 93551R and HI 93552R models are equipped with a serial port interface WARRANTY...

Page 3: ...to 0 1 F 0 3 F 328 0 to 270 1 F Accuracy 0 5 C 100 0 to 999 9 C 1 C outside 20 C 68 F 1 F 148 0 to 999 9 F 1 5 F outside for one year excluding probe error Typical EMC Deviation 3 C 6 F Battery 3x1 5V...

Page 4: ...t the thermocouple type ALT BL toggle the backlight ON and OFF SPECIFICATIONS Range K 200 0 to 999 9 C 1000 to 1371 C 328 0 to 999 9 F 1000 to 2500 F J 200 0 to 999 9 C 328 0 to 999 9 F 1000 to 1832 F...

Page 5: ...ght ON and OFF SPECIFICATIONS Range K 200 0 to 999 9 C 1000 to 1371 C 328 0 to 999 9 F 1000 to 2500 F J 200 0 to 999 9 C 328 0 to 999 9 F 1000 to 1832 F T 200 0 to 400 0 C 328 0 to 752 0 F Resolution...

Page 6: ...this feature set the internal slide switch located in the battery compart ment see figure on page 10 OPERATIONAL GUIDE To switch the instrument ON press the ON OFF key The thermometer will carry out a...

Page 7: ...ags light up to inform the user To select the desired channel use the T1 T2 key The 2 channel and HI 93551R models are also provided with the DISP function which allows the user to select the informat...

Page 8: ...HI 93551R and HI 93552R feature an RS232 output for transfer ring measurement data once every second for HI 93551R and once every 2 seconds for HI 93552R to devices provided with RS232 input PC or pr...

Page 9: ...your nearest Hanna Customer Service Center When the battery level is below 5 a warning symbol will blink on the LCD to indicate a low battery condition If the battery level is low enough to cause err...

Page 10: ...n long probe max 250 C 482 F HI 766TV1 Pipe clamp probe max 200 C 390 F with detachable handle mini connector to be used in conjunction with the HI 766HD probe handle HI 766PA Roller surface probe max...

Page 11: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax...

Reviews: