background image

Zero Calibration:

• Dip the probe into HI 7040 zero

oxygen solution and stir gen-
tly for 2-3 minutes.

• Allow 2 minutes for the LCD readout to

stabilize.

• Adjust the zero D.O. calibration

trimmer until the display reads
"0.0".

Note  The zero calibration screw is
located close to the probe
connector.

Slope Calibration:

It is suggested to perform the slope calibration
in satured air.

• Rinse the probe in a large amount of clean

water to remove any residual zero oxygen
solution.

• Dry the probe tip and allow a few minutes for

the LCD readout to stabilize.

• Press and hold the CAL key.

• Adjust the slope trimmer on

the top of the meter to read
"100%" on the LCD (while still
holding the CAL button).

Note The slope calibration screw

is located farthest from the probe
connection.

• Release the CAL key and the

LCD will display the value in
ppm of oxygen.

The zero calibration of the HI 9142 is very
stable, therefore this procedure needs only to
be performed whenever the probe is re-
placed
.

However, because the slope calibration is
more critical, it is recommended to perform
this procedure every week
.

%

CAL

7

CALIBRATION  PROCEDURE

PROBE   POLARIZATION
The probe is under polarization with a fixed
voltage of approximately 800 mV.
Probe polarization is essential for stable mea-
surements with the same recurring degree of
accuracy.
With the probe properly polarized, oxygen is
continually "consumed" by passing through the
sensitive diaphragm and dissolving in the elec-
trolyte solution contained in the probe.
If this operation is interrupted, the electrolyte
solution continues to be enriched with oxygen
until it reaches an equilibrium with the surround-
ing solution.
Whenever measurements are taken with a non-
polarized probe, the oxygen level revealed is
both that of the tested solution as well as that
present in the electrolyte solution. This reading
is incorrect

The calibration is very simple and fast.

• Make sure the probe is ready for measure-

ments (see initial preparation at page 7), i.e.
the membrane is filled with electrolyte and the
probe is connected to the meter.

• Switch the meter on by press-

ing the ON/OFF key

• For an accurate calibration, it is recommended

that you wait at least 15 minutes to ensure
precise conditioning of the probe.

• Remove the protective cap

from the D.O. probe.

ON/OFF

6

Summary of Contents for HI 9142

Page 1: ...Instruction Manual HI 9142 Portable Waterproof Dissolved Oxygen Meter This instrument is in Compliance with the CE Directives http www hannainst com ...

Page 2: ...s per million ppm mg l Four1 5VAAbatteriesprovidepowerandmake theinstrumentcompletelyportable Themeterishousedinaruggedwater resistant caseformaximumprotectioninthefieldaswell as in the laboratory The dissolved oxygen probe has a membrane coveringthepolarographicsensorsandabuilt 1 DearCustomer Thank you for choosing a Hanna Instruments Product Please read this instruction manual carefully before u...

Page 3: ...or elements from the testing solution but allows oxygen to enter When a voltage is applied across the sensor oxygenthathaspassedthroughthemembrane reacts causing current to flow allowing the determinationofoxygen 1 D O Probe 2 ProtectiveCap 3 Water tight Shielded Cable 4 Polypropylene Probe Body 5 TemperatureSensor 6 O RingSeal 7 Silver Chloride Anode 8 Platinum Cathode sensor 9 OxygenPermeablePTF...

Page 4: ...aging the mem brane do not tap the membrane directly on the bottom 5 Place rubber O Ring properly inside the membrane cap 6 With the sensor facing down screw the cap clockwise Some electrolyte will overflow When not in use and during polar ization place the protective cap supplied in the kit with the meter Shipping cap black red THEN SCREW BACK ON THEN TAP FILL FIRST 5 HI9142 Range 0 0 to 19 9 mg ...

Page 5: ...t is recommended to perform this procedure every week CAL 7 CALIBRATION PROCEDURE PROBE POLARIZATION The probe is under polarization with a fixed voltage of approximately 800 mV Probe polarization is essential for stable mea surements with the same recurring degree of accuracy With the probe properly polarized oxygen is continually consumed bypassingthroughthe sensitivediaphragmanddissolvinginthee...

Page 6: ...5 9 5 8 Altitude Meters above Sea Level This gives an idea of the error that can be introducedatdifferentaltitudesandallowsyou to calculate the quantity to be subtracted to correct your reading F 32 0 35 6 39 2 42 8 46 4 50 0 53 6 57 2 60 8 64 4 68 0 71 6 75 2 78 8 82 4 86 0 89 6 93 2 96 8 100 4 104 4 9 TAKING MEASUREMENTS Makesurethemeterhasbeen calibrated and the protective cap has been removed ...

Page 7: ...esidesofthe membrane with your fin ger tip to ensure that no air bubbles remain trapped Do no directly tap the bottom with your fingerasthiswilldamage themembrane TWIST AND PULL fig 1 UNSCREW fig 2 11 SALINITY COMPENSATION The table below shows the influence of salt in the measurement of oxygen In HI 9142 all the readouts are referred to 0 g l of salinity value In fact salinity affects D O concent...

Page 8: ...ription at page 4 should always be bright and untarnished If it is tarnished or stained which could be due to contact with certain gasesorextendedusewithalooseordamaged membrane thecathodeshouldbecleaned You canuseacleanlint freecardboardorcloth Rub the cathode very gently side to side 4 5 times This will be enough to polish and remove any stainswithoutdamagingtheplatinumtip After wards rinsethepr...

Page 9: ...nedtoHannaInstruments obtainaReturn Goods Authorization from the Customer Ser vice Department first and then send it with shipment cost prepaid When shipping any instrument make sure it is properly packaged for complete protection Tovalidateyourwarranty filloutandreturnthe enclosedwarrantycardwithin14daysfromthe date of purchase All rights are reserved Reproduction in whole or in part is prohibite...

Page 10: ...vens In particular cases the meter could turn off In these cases it can be turned on by pressing the ON OFF key 17 CALIBRATION AND MAINTENANCE SOLUTIONS CHEMICAL TEST KITS CHLORINE METERS CONDUCTIVITY TDS METERS DISSOLVED OXYGEN METERS HYGROMETERS ION SPECIFIC METERS Colorimeters MAGNETIC STIRRERS pH Na NaCl METERS pH ORP Na ELECTRODES PROBES DO µS cm RH T TDS PUMPS REAGENTS SOFTWARE THERMOMETERS ...

Page 11: ...MAN9142R1 12 95 PRINTED IN PORTUGAL h t t p w w w h a n n a i n s t c o m ...

Reviews: