background image

ALTITUDE  &  SALINITY  COMPENSATION

If the sample contains salts or if you are
performing the measurements at a different
altitude than sea level, the readout values must
be corrected, taking into account the lower
degree of oxygen solubility.

ALTITUDE  COMPENSATION

All the readouts are referred at sea level, thus
the displayed measurements are higher than
the actual values. In fact, altitude affects D.O.
concentration decreasing its value. The follow-
ing table reports the oxygen solubility at various
temperatures and altitudes, based on sea level
barometric pressure of 760 mm Hg.

°C

0
2
4
6
8

10
12
14
16
18
20
22
24
26
28
30
32
34
36
38
40

0 m

14.6
13.8
13.1
12.4
11.8
11.3
10.8
10.3

9.9
9.5
9.1
8.7
8.4
8.1
7.8
7.5
7.3
7.1
6.8
6.6
6.4

300 m

14.1
13.3
12.7
12.0
11.4
10.9
10.4

9.9
9.7
9.2
8.8
8.4
8.1
7.8
7.5
7.2
7.1
6.9
6.6
6.4
6.2

600 m

13.6
12.9
12.2
11.6
11.0
10.5
10.1

9.6
9.2
8.7
8.5
8.1
7.8
7.5
7.3
7.0
6.8
6.6

   6.3
   6.2
   6.0

1200 m

12.7
12.0
11.4
10.8

10.3

9.8
9.4
9.0
8.6
8.3
7.9
7.7
7.3
7.0
6.8
6.5

      6.4
      6.2
      5.9
      5.7
      5.6

1500m

12.3
11.6
11.0

10.4

9.9
9.5
9.1
8.7
8.3
8.0
7.7
7.3
7.1
6.8
6.6

      6.3
      6.1
      6.0
      5.7
      5.6
      5.4

1800m

11.8
11.2
10.6
10.1

9.6
9.2
8.8
8.3
8.0
7.7
7.4
7.1
6.8
6.6

    6.3
    6.1
    5.9
    5.8
    5.5
    5.4
    5.2

900 m

13.2
12.4
11.9
11.2
10.6
10.2

9.7
9.3
8.9
8.6
8.2
7.8
7.5
7.3
7.0
6.8
6.6

      6.4
      6.1
      5.9
      5.8

Altitude, Meters above Sea Level

This gives an idea of the error that can be
introduced at different altitudes and allows you
to calculate the quantity to be subtracted to
correct your reading.

°F

32.0
35.6
39.2
42.8
46.4
50.0
53.6
57.2
60.8
64.4
68.0
71.6
75.2
78.8
82.4
86.0
89.6
93.2
96.8

100.4
104.4

9

TAKING   MEASUREMENTS

Make sure the meter has been
calibrated and the protective
cap has been removed. Im-
merse the tip of the probe in the
sample to be tested.

Make sure the temperature sensor is also
immersed.

For accurate dissolved oxygen measurements
a water movement of 0.3 m/sec is required at a
minimum. This is to ensure that the oxygen-
depleted membrane surface is constantly re-
plenished. A moving stream will provide ad-
equate circulation. To quickly check if the water
speed is sufficient, wait for the reading to
stabilize and then move the D.O. probe. If the
reading is still stable, the measurement condi-
tions are right, while if the reading increases the
water movement is not adequate.

During field measurements, this condition may
be met by manually agitating the probe. Accu-
rate readings are not possible while the liquid is
at rest.

During laboratory measurements, the use of  a
magnetic stirrer to ensure a certain velocity in
the fluid is recommended. In this way, errors
due to the diffusion of the oxygen present in the
air in the solution  are reduced to a minimum.

At all times, time necessary for thermal equilib-
rium to occur between the probe and the sample
must be allowed (a few minutes for temperature
difference of several degrees).

8

Summary of Contents for HI 9142

Page 1: ...Instruction Manual HI 9142 Portable Waterproof Dissolved Oxygen Meter This instrument is in Compliance with the CE Directives http www hannainst com ...

Page 2: ...s per million ppm mg l Four1 5VAAbatteriesprovidepowerandmake theinstrumentcompletelyportable Themeterishousedinaruggedwater resistant caseformaximumprotectioninthefieldaswell as in the laboratory The dissolved oxygen probe has a membrane coveringthepolarographicsensorsandabuilt 1 DearCustomer Thank you for choosing a Hanna Instruments Product Please read this instruction manual carefully before u...

Page 3: ...or elements from the testing solution but allows oxygen to enter When a voltage is applied across the sensor oxygenthathaspassedthroughthemembrane reacts causing current to flow allowing the determinationofoxygen 1 D O Probe 2 ProtectiveCap 3 Water tight Shielded Cable 4 Polypropylene Probe Body 5 TemperatureSensor 6 O RingSeal 7 Silver Chloride Anode 8 Platinum Cathode sensor 9 OxygenPermeablePTF...

Page 4: ...aging the mem brane do not tap the membrane directly on the bottom 5 Place rubber O Ring properly inside the membrane cap 6 With the sensor facing down screw the cap clockwise Some electrolyte will overflow When not in use and during polar ization place the protective cap supplied in the kit with the meter Shipping cap black red THEN SCREW BACK ON THEN TAP FILL FIRST 5 HI9142 Range 0 0 to 19 9 mg ...

Page 5: ...t is recommended to perform this procedure every week CAL 7 CALIBRATION PROCEDURE PROBE POLARIZATION The probe is under polarization with a fixed voltage of approximately 800 mV Probe polarization is essential for stable mea surements with the same recurring degree of accuracy With the probe properly polarized oxygen is continually consumed bypassingthroughthe sensitivediaphragmanddissolvinginthee...

Page 6: ...5 9 5 8 Altitude Meters above Sea Level This gives an idea of the error that can be introducedatdifferentaltitudesandallowsyou to calculate the quantity to be subtracted to correct your reading F 32 0 35 6 39 2 42 8 46 4 50 0 53 6 57 2 60 8 64 4 68 0 71 6 75 2 78 8 82 4 86 0 89 6 93 2 96 8 100 4 104 4 9 TAKING MEASUREMENTS Makesurethemeterhasbeen calibrated and the protective cap has been removed ...

Page 7: ...esidesofthe membrane with your fin ger tip to ensure that no air bubbles remain trapped Do no directly tap the bottom with your fingerasthiswilldamage themembrane TWIST AND PULL fig 1 UNSCREW fig 2 11 SALINITY COMPENSATION The table below shows the influence of salt in the measurement of oxygen In HI 9142 all the readouts are referred to 0 g l of salinity value In fact salinity affects D O concent...

Page 8: ...ription at page 4 should always be bright and untarnished If it is tarnished or stained which could be due to contact with certain gasesorextendedusewithalooseordamaged membrane thecathodeshouldbecleaned You canuseacleanlint freecardboardorcloth Rub the cathode very gently side to side 4 5 times This will be enough to polish and remove any stainswithoutdamagingtheplatinumtip After wards rinsethepr...

Page 9: ...nedtoHannaInstruments obtainaReturn Goods Authorization from the Customer Ser vice Department first and then send it with shipment cost prepaid When shipping any instrument make sure it is properly packaged for complete protection Tovalidateyourwarranty filloutandreturnthe enclosedwarrantycardwithin14daysfromthe date of purchase All rights are reserved Reproduction in whole or in part is prohibite...

Page 10: ...vens In particular cases the meter could turn off In these cases it can be turned on by pressing the ON OFF key 17 CALIBRATION AND MAINTENANCE SOLUTIONS CHEMICAL TEST KITS CHLORINE METERS CONDUCTIVITY TDS METERS DISSOLVED OXYGEN METERS HYGROMETERS ION SPECIFIC METERS Colorimeters MAGNETIC STIRRERS pH Na NaCl METERS pH ORP Na ELECTRODES PROBES DO µS cm RH T TDS PUMPS REAGENTS SOFTWARE THERMOMETERS ...

Page 11: ...MAN9142R1 12 95 PRINTED IN PORTUGAL h t t p w w w h a n n a i n s t c o m ...

Reviews: