FUNCTIONAL DESCRIPTION HI8014
ins rum n s
pH
pH meter
HI 8014
°C
mV/pH
AP
pH
°C
OFF
mV
2
1
3
4
1) BNC socket for combination pH electrodes.
2) Liquid Crystal Display (LCD).
3)
Rotary Knob
with the following selectable positions:
OFF
To switch the meter off (in any other position the
instrument is on).
°C
To display the manual temperature setting (through
°C trimmer).
pH
To display the pH value.
mV
To display the mV (ORP) readings when using an
ORP electrode or the mV equivalent to the pH values
when using a pH electrode.
4)
Trimmers:
°C
To set temperature manually.
mV/pH
For SLOPE calibration of pH.
AP
For OFFSET calibration of pH.
SPECIFICATIONS HI8014
HI 8014
Range
pH
0.00 to 14.00
mV
±1999
Resolution
pH
0.01
mV
1
Accuracy
pH
±0.01
(@20°C/68°F)
mV
±1
Typical EMC
pH
±0.05
Deviation
mV
±8
pH Calibration
Manual 2 point through
offset and slope trimmers
Offset Calibration
±1 pH
Slope Calibration
from 85 to 105%
Temperature
Manual from 0 to 100°C
Compensation
(32 to 212°F)
pH Electrode
HI 1230B
combination, double-junction,
gel-filled, with BNC connector
and 1 m (3.3') cable (
included
)
Input Impedance
10
12
ohm
Battery Type
1 x 9V
Life
100 hours of continuous use
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Dimensions
185 x 82 x 53 mm
(7.3 x 3.2 x 2.1’’)
Weight
265 g (9.3 oz.)