5

Light Detector

Silicon Photocell

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

4 x 1,5 volt AA batteries / 12 to 20 VDC through voltage adapter

Auto-Shut off

After 15' of non-use.

Dimensions

225 x 85 x 80 mm (8.7 x 3.3 x 3.1")

Weight

500 g (17,6 oz.).

SPECIFICATIONS

Precision is how closely repeated measurements agree
with each other.  Precision is usually expressed as
standard deviation (SD). Accuracy is defined as the
nearness of a test result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure explains
these definitions.
For example for phenols: in a laboratory using a
standard solution of 0.350 g/L gallic acid and a
representative lot of reagent, an operator obtained
with a single instrument a standard deviation of

0.015 g/L.

PRECISION AND ACCURACY

The reaction between phenols and the Folin-Ciocalteu reagent involves oxidation of the phenolic groups (R-
OH) with a mixture of phosphotungstenic acid (H

3

PW

12

O

40

) and phosphomolybdenic acid (H

3

PMo

12

O

40

) to

the quinoid form (R=O). The concomitant reduction of the Folin-Ciocalteu reagent causes a blue color in
the sample that is proportional to the total phenolic content that, in turn, is expressed as g/L of Gallic Acid
Equivalents (GAE).
Another common method to express the phenol content is to report the Folin Index (FI) or Total Phenol
Index. Simply multiply the read value in g/L of GAE times 25, to find the Folin Index value.

24

• Add 6 drops of HI 83742B-0 Folin & Ciocalteu’s

Reagent to the cuvet.

• Replace the cap and shake gently to mix.

x 6

#2

(Sample)

• Use the 200 µL automatic pipette with a clean

pipette tip to add exactly 0,2 mL of 

diluted red

wine sample

.

10 mL

• Wait for 1 minute and then use the 3 mL pipette to

add HI 83742C-0 Reagent C Carbonate Buffer, to
bring the volume up to the 10 mL mark.

• Turn the meter ON and press RANGE to select the

parameter code “P5” for total phenols, red wine (see
page 13). Then hold down READ/TIMER for several
seconds. The instrument will show the countdown.
Alternatively, wait for 2 hours.
The instrument gives an acoustic signal to alert the
user that the countdown is finished.

• This is the reacted sample.

Summary of Contents for HI 83742

Page 1: ...Instruction Manual HI 83742 COLOR PHENOLS ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83742R3 04 06 w w w h a n n a i n s t c o m...

Page 2: ...tampering or lack of prescribed maintenance are not covered If service is required contact your dealer If under warranty report the model number date of purchase serial number and the nature of the f...

Page 3: ...hazardous area The blinking will appear when the batteries power gets low When batteries are completely discharged 0 bAtt will appear and after two seconds the instrument is switched off Remove the b...

Page 4: ...ecommends usingthespecialwinesolventtominimizepossibleerrorsduetodilution Phenoldeterminationofwine Phenoliccompoundsareimportantforseveralreasonssincethey i affectthecolorofthewine ii have anastringe...

Page 5: ...ungstenicacid H3PW12O40 andphosphomolybdenicacid H3PMo12O40 to thequinoidform R O TheconcomitantreductionoftheFolin Ciocalteureagentcausesabluecolorin thesamplethatisproportionaltothetotalphenoliccont...

Page 6: ...s colorimeters is based on special subminiature tungsten lamps and narrow band interference filters to guarantee both high performance and reliable results Block diagram optical layout 23 TOTAL PHENOL...

Page 7: ...This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions...

Page 8: ...to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 21 Add 6 drops of HI 83742B 0 Folin Ciocalteu s Reagent to the cuvet Replace the cap and shake gently to mix x 6 Wait for 1...

Page 9: ...replaced 20 TOTAL PHENOLS WHITE WINE SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Range 0 000 to 0 750 g L Resolution 0 001 Precision 0 015 0 350 g L Light Source Tungste...

Page 10: ...e cuvet for reference zero and measurement Inverted the sample and the zero cuvet are inverted b on sample reading b on sample reading b on sample reading b on sample reading b on sample reading Indic...

Page 11: ...hen keeping the syringe in vertical position above the cuvet push the plunger completely down into the syringe Now the exact amount of 5 mL has been added to the cuvet GENERAL TIPS FOR AN ACCURATE MEA...

Page 12: ...nual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F 17 MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT Press RANGE to sele...

Page 13: ...nm Method Direct reading REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS Code Description Quantity test HI 83742 0 Wine Solvent 1 8 mL HI 83742 3 Wine Solven...

Page 14: ...the zero zero zero zero zero Insert the zero cuvet into the holder and close the lid Press ZERO and will blink on the display SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATIO...