MEASUREMENT PROCEDURE

OPERATIONAL GUIDE

• Turn the instrument on by pressing ON/OFF.

PARAMETER REFERENCE TABLE

• After the desired parameter code appears on the display,

perform a measurement procedure following the
instructions reported in the related chapter.

HI 83742 - COLOR & PHENOLS

Code

Parameter

Page

1

Color Density, white wine

14

2

Color Density, red wine

16

3

Color Hue

18

4

Total Phenols, white wine

20

5

Total Phenols, red wine

23

The meter will first perform an LCD self diagnostic test
by displaying a full set of figures.

• When the LCD displays “----”, it is ready. On the

secondary LCD a code indicates which parameter is
selected (“P1” or “P2” or... “P5”). The display code
that appears is the one of the last selected parameter.

Press the RANGE key to select the desired parameter.

16

COLOR DENSITY, RED WINE

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS
Range

0.00 to 15.00

Resolution

0.01

Precision

±0.20 @ 5.00

Light Source

Tungsten lamp with narrow band interference filters @ 420 and 520 nm

Method

Direct reading.

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 83742-0

Wine Solvent-1

8 mL

HI 83742-3

Wine Solvent-3

8 mL [optional reagent]

For reagent set and other accessories see page 26.

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

• Replace the cap. This is the sample.

#2

(Sample)

• Use the 2000 µL automatic pipette to add exactly

2 mL of red wine sample to an empty cuvet. For a
correct use of the automatic pipette please follow the
related Instruction Sheet.

• Then use the 1 mL plastic pipette to fill the cuvet

with HI 83742-0 Wine Solvent-1 (or with HI 83742-3
Wine Solvent-3 - optional reagent) up to the 10 mL
mark.
Note: HI 83742-0 Wine Solvent 1 is used to measure

the true color of red wines and the results obtained
are conform to EC Directive 2676/90. The
corresponding replacement kit is HI 83742-25
(see page 26).

Note: HI 83742-3 Wine Solvent 3 is a stabilized and

ready-to-use solution at a fixed pH value (=3.2)
to compare different types of dark colored wine
samples often used for blending. The corresponding
replacement kit is HI 83742-27 (see page 26).

10 mL

(          )

optional

or

Summary of Contents for HI 83742

Page 1: ...Instruction Manual HI 83742 COLOR PHENOLS ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83742R3 04 06 w w w h a n n a i n s t c o m...

Page 2: ...tampering or lack of prescribed maintenance are not covered If service is required contact your dealer If under warranty report the model number date of purchase serial number and the nature of the f...

Page 3: ...hazardous area The blinking will appear when the batteries power gets low When batteries are completely discharged 0 bAtt will appear and after two seconds the instrument is switched off Remove the b...

Page 4: ...ecommends usingthespecialwinesolventtominimizepossibleerrorsduetodilution Phenoldeterminationofwine Phenoliccompoundsareimportantforseveralreasonssincethey i affectthecolorofthewine ii have anastringe...

Page 5: ...ungstenicacid H3PW12O40 andphosphomolybdenicacid H3PMo12O40 to thequinoidform R O TheconcomitantreductionoftheFolin Ciocalteureagentcausesabluecolorin thesamplethatisproportionaltothetotalphenoliccont...

Page 6: ...s colorimeters is based on special subminiature tungsten lamps and narrow band interference filters to guarantee both high performance and reliable results Block diagram optical layout 23 TOTAL PHENOL...

Page 7: ...This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions...

Page 8: ...to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 21 Add 6 drops of HI 83742B 0 Folin Ciocalteu s Reagent to the cuvet Replace the cap and shake gently to mix x 6 Wait for 1...

Page 9: ...replaced 20 TOTAL PHENOLS WHITE WINE SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Range 0 000 to 0 750 g L Resolution 0 001 Precision 0 015 0 350 g L Light Source Tungste...

Page 10: ...e cuvet for reference zero and measurement Inverted the sample and the zero cuvet are inverted b on sample reading b on sample reading b on sample reading b on sample reading b on sample reading Indic...

Page 11: ...hen keeping the syringe in vertical position above the cuvet push the plunger completely down into the syringe Now the exact amount of 5 mL has been added to the cuvet GENERAL TIPS FOR AN ACCURATE MEA...

Page 12: ...nual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F 17 MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT Press RANGE to sele...

Page 13: ...nm Method Direct reading REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS Code Description Quantity test HI 83742 0 Wine Solvent 1 8 mL HI 83742 3 Wine Solven...

Page 14: ...the zero zero zero zero zero Insert the zero cuvet into the holder and close the lid Press ZERO and will blink on the display SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATIO...

Reviews: