
13
12
HI 504910 pH READING CHECK
The check of the
HI 504910
pH reading against the
HI 504920
pH
reading is possible at any time.
Perform the following steps:
• Connect the infrared optical interface cable to
HI 504910
.
• Press (ALT+) TXCAL.
• The primary LCD shows the
HI 504910
pH reading and the secondary one dis-
plays the
HI 504920
pH reading. Both
pH tags blink alternatively.
• Press (ALT+) TXCAL to return to normal mode.
Note:
In case of a communication error
between
HI 504920
and
HI 504910
,
the LCD shows the "rSEr" message.
Check interface cable and connections.
Note:
If the LCD shows "...", verify that
HI 504910
is configured to
measure pH and not ORP.
Note:
At point 5, if the probe offset is
outside the allowed limits, if
HI
504910
is configured to measure ORP,
or if some other error occurs, the LCD
shows a blinking "bAd" message. To cancel this message and
return to normal mode, press TXCAL. Then restart the calibra-
tion procedure from the beginning.
Note:
Although it is recommended to perform the entire calibration
procedure with the infrared optical interface cable connected
between
HI 504920
and
HI 504910
(step 1), it can happen
that the
HI 504910
must be located in a place difficult to
reach when the
HI 504920
pH probe is immersed in the
solution.
In this case, step 1 can be skipped. After step 4 the primary
LCD will show the blinking "CALt" message and the secondary
one a countdown starting from 120 seconds.
In this situation, the pH probe can be taken out of the
solution and the interface cable must be connected to the
HI
504910
before the countdown reaches zero.
Calibration will then continue from step 5. At step 6 the pH
tag of the secondary LCD does not blink, because the corre-
sponding pH reading is the one taken with the
HI 504920
pH
probe immersed when TXCAL was pressed.
If the interface cable is not connected before the end of the
countdown, the LCD shows
the "tiMEout" message and
the calibration procedure is
aborted.
Note:
To perform a two-point calibration procedure with standard pH
buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) through the
HI
504910
terminal mode, see the "Setup mode section", code
90.
Note:
A probe life verification (as described in the "GLP" section) is
performed on the
HI 504910
pH probe; the message "old
probe" or "dead probe" could appear at the end of calibration.