Hanna Instruments HI 504920 Instruction Manual Download Page 5

9

8

CALIBRATION

pH CALIBRATION

For greatest accuracy, it is recommended to frequently calibrate the
instrument. For a faster procedure, it is possible to calibrate at 1
point, but it is always a good practice to calibrate at 2 points.
For a two-point calibration two of the following pH buffers can be
used: pH 4.01, pH 7.01 (or 6.86), pH 10.01 (or 9.18).
In the case of a two-point calibration in the acidic range (from 0 to
7 pH), use pH 7.01 (or 6.86) buffer as first solution and pH 4.01
buffer as second solution. If testing in the alkaline range (from 7 to
14 pH), use pH 10.01 (or 9.18) buffer as second solution.
Due to electrode conditioning time, the electrode must be kept
immersed a few seconds to stabilize. The meter is equipped with a
stability indicator and the user will be guided step by step with easy
indications on the LCD during the calibration. This will make the
calibration a simple and error-free procedure.

PROCEDURE

1. Rinse the electrode with a portion of the first  calibration buffer or

clean water. Dip the tip (4 cm/1½") of the electrode into a
beaker containing the solution.

2. Press (ALT+) CAL when the meter dis-

plays pH measurement.

3. Enter the password (if different from "0000") with the arrow keys.
4. Press CFM to confirm the password or (ALT+) CAL to exit.
5. If password is correct, the meter displays "7.01pH" on the

secondary LCD with the "BUF 1" and "CAL" indications. The
primary LCD shows the uncalibrated pH reading.

Note

: The pH buffer value (displayed on the secondary LCD) varies

with temperature. For example, at 20

°

C it shows 4.00 -7.03 -

10.06, at 25

°

C it shows 4.01-7.01-10.01.

6. Select the first buffer solution value with 

 or 

 if necessary.

7. When the "CFM" symbol blinks, the read-

ing is stable and calibration can be
confirmed.

8. Press CFM to confirm the first buffer.

9. If everything is satisfactory the LCD will

display "Stor" and then the expected
value for the second buffer (two-point
calibration). If a wrong solution or elec-
trode has been used or if the buffer is
polluted, "WRONG" will be displayed to
alert the user.

10. If a single point calibration is required,

press (ALT+) CAL to exit the calibration
mode and maintain the previous slope
calibration. The instrument then checks
the electrode parameters and advises user
of possible abnormalities by "old probe"
and "dead probe" indications.

11. Press 

 or 

 to select the second buffer value. The meter will

display the "BUF 2" indication.

12. Rinse the electrode with some of the second buffer solution or clean

water.

13. Dip the tip (4 cm/1½") of the pH electrode into a beaker

containing the second buffer.

14. When the "CFM" symbol blinks, press CFM

to confirm the second calibration point.

15. The LCD will display "Stor". The instrument checks the electrode

parameters and advises the user of possible abnormalities by "old
probe" and "dead probe" indications (in these cases, repeat the
calibration procedure with fresh buffers). If everything is satisfactory
the meter is calibrated and it returns to normal operational mode.

Note

: To abort the calibration procedure and leave the calibration data

unchanged, press ALT and CAL before confirming the first point.

CALIBRATION ERROR MESSAGES

If the "old probe" or "dead probe" messages are displayed, or if the
WRONG tag blinks during calibration, check the electrode by following
the conditioning and maintenance procedures and repeat calibration.
If calibration can not be successfully performed, the pH electrode
might have to be replaced.

Note

: See "GLP" section for more details about "old probe" and "dead

probe" messages.

Summary of Contents for HI 504920

Page 1: ...Instruction Manual HI 504920 Remote User Interface pH Meter w w w h a n n a i n s t c o m...

Page 2: ...PRELIMINARY EXAMINATION GENERAL DESCRIPTION Dear Customer Thank you for choosing a Hanna Instruments Product Please read this instruction manual carefully before using the instru ment This manual wil...

Page 3: ...Preventing System BEPS detects when the batter ies level becomes weak When the batteries are getting low a clear indication is displayed on LCD to warn the user of this condition However the meter co...

Page 4: ...efore turning the meter off When the meter is OFF detach the electrode from the meter before immersing it in the storage solution SPECIFICATIONS Range pH 4 00 to 19 99 pH mV 600 0 mV 2000 mV autorangi...

Page 5: ...When the CFM symbol blinks the read ing is stable and calibration can be confirmed 8 Press CFM to confirm the first buffer 9 If everything is satisfactory the LCD will display Stor and then the expect...

Page 6: ...e above procedure only when readings are stable 7 To exit the calibration mode and return to normal measurement press TXCAL mV CALIBRATION Contact your nearest Hanna Service Center for periodic mV cal...

Page 7: ...nterface cable connected between HI 504920 and HI 504910 step 1 it can happen that the HI 504910 must be located in a place difficult to reach when the HI 504920 pH probe is immersed in the solution I...

Page 8: ...0 Calibration alarm time out 01 to 99 days OFF OFF 20 Auto Off Power down time out ON OFF ON 21 Firmware version 22 Battery level test 90 HI 504910 terminal mode 99 Password2 0000 to 9999 0000 Setup c...

Page 9: ...y checks for entered time date accuracy as follows 0 hh 23 0 mm 59 01 dd 28 29 30 31 1 MM 12 1998 YYYY 2097 2 To change the password the correct code must be entered first If the password has been for...

Page 10: ...When turned on the meter checks if the time out time has expired If yes the message Cal date scrolls across the LCD and the DATE tag blinks as a reminder LAST CALIBRATION DATA Last calibration data ar...

Page 11: ...ers In this case check the correct polarity of the adapter before connecting it to the meter BATTERY REPLACEMENT The LCD will then display the last pH calibration date Press DISPLAY to scan remaining...

Page 12: ...4 00 12 50 13 00 13 50 14 00 0 10 0 14 0 20 0 10 0 18 0 29 0 40 Note To prevent damage to the electrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach...

Page 13: ...torage Solution Follow the Preparation Procedure above before taking measurements Note NEVER STORE THE ELECTRODE IN DISTILLED WATER OR DRY PERIODIC MAINTENANCE Inspect the electrode and the cable The...

Page 14: ...SOLUTIONS HI 7091L Reducing Pretreatment solution 500 mL bottle HI 7092L Oxidizing Pretreatment solution 500 mL bottle ELECTRODES HI 62920 pH electrode with built in temperature sensor matching pin Ti...

Page 15: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Reviews: