Hanna Instruments HI 504920 Instruction Manual Download Page 13

25

24

ACCESSORIES

pH CALIBRATION SOLUTIONS

HI 70004P

pH 4.01 buffer, 20 mL sachet (25 pcs)

HI 70007P

pH 7.01 buffer, 20 mL sachet (25 pcs)

HI 70010P

pH 10.01 buffer, 20 mL sachet (25 pcs)

HI 7004L

pH 4.01 buffer, 500 mL bottle

HI 7006L

pH 6.86 buffer, 500 mL bottle

HI 7007L

pH 7.01 buffer, 500 mL bottle

HI 7009L

pH 9.18 buffer, 500 mL bottle

HI 7010L

pH 10.01 buffer, 500 mL bottle

ELECTRODE STORAGE SOLUTION

HI 70300L

Storage solution, 500 mL bottle

ELECTRODE CLEANING SOLUTIONS

HI 70000P

Rinsing solution, 20 mL sachet (25 pcs)

HI 7061L

General cleaning solution, 500 mL bottle

HI 7073L

Protein cleaning solution, 500 mL bottle

HI 7074L

Inorganic cleaning solution, 500 mL bottle

HI 7077L

Oil & Fat cleaning solution, 500 mL bottle

REFILL ELECTROLYTE SOLUTIONS

HI 7071

3.5M KCl + AgCl Electrolyte, 50 mL (4 pcs)

HI 7072

1M KNO

3

 Electrolyte, 50 mL (4 pcs)

HI 7082

3.5M KCl Electrolyte, for double junction electrodes, 50
mL (4 pcs)

• Dry Membrane/Junction:

 Soak in 

Storage Solution HI 70300

for at least 1 hour.

• Drifting:

 Soak the electrode tip in warm Hanna Solution 

HI

7082

 for 1 hour, then flush tip with distilled water. Refill with

fresh 

HI 7071

 for single junction electrodes and 

HI 7082 

for

double junction electrodes.

• Low Slope:

 Refer to the cleaning procedure above.

• No Slope: 

Check the electrode for cracks in glass stem or bulb

and replace the electrode.

• Slow Response/Excessive Drift: 

Soak the tip in Hanna Solu-

tion 

HI 7061 

for 30 minutes, rinse thoroughly in distilled water

and then follow the Cleaning Procedure above.

Replace the solution in the protective cap with a few drops of 

HI

70300 Storage Solution

. Follow the Preparation Procedure above

before taking measurements.

Note:

NEVER STORE THE ELECTRODE IN DISTILLED WATER OR DRY.

PERIODIC MAINTENANCE

Inspect the electrode and the cable. The cable used for connection to
the meter must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb.
Connectors must be perfectly clean and dry.
If any scratches or cracks are present, replace the electrode.
Rinse off any salt deposits with water.

For refillable electrodes:

Refill it with fresh electrolyte (

HI 7071

 for single junction or 

HI 7082

for double junction electrodes). Allow the electrode to stand upright
for 1 hour.
Follow the Storage Procedure above.

CLEANING PROCEDURE

General Soak in Hanna 

HI7061

 

General Cleaning Solution

 for

approximately 1 hour.
Removal of films, dirt or deposits on the membrane/junction:

Protein

Soak in Hanna 

HI 7073

 

Protein Cleaning

Solution

 for 15 minutes.

Inorganic

Soak in Hanna 

HI 7074 Inorganic Cleaning

Solution

 for 15 minutes.

Oil/grease

Rinse with Hanna 

HI 7077 Oil and Fat Clean

ing Solution

.

IMPORTANT:

 After performing any of the cleaning procedures rinse

the electrode thoroughly with distilled water, drain and refill the
reference chamber with fresh electrolyte, (not necessary for GEL filled
electrodes) and soak the electrode in 

HI 70300

 

Storage Solution

 for

at least 1 hour before taking measurements.

TROUBLESHOOTING

Evaluate your electrode performance based on the following possibilities:

• Noise 

(Readings fluctuate up and down) could be due to:

- Clogged/Dirty Junction: 

Refer to the Cleaning Procedure above.

- Loss of shielding

 due to low electrolyte level (in refillable

electrodes only): refill with fresh 

HI 7071 

for single junction or 

HI

7082 

for double junction electrodes.

Summary of Contents for HI 504920

Page 1: ...Instruction Manual HI 504920 Remote User Interface pH Meter w w w h a n n a i n s t c o m...

Page 2: ...PRELIMINARY EXAMINATION GENERAL DESCRIPTION Dear Customer Thank you for choosing a Hanna Instruments Product Please read this instruction manual carefully before using the instru ment This manual wil...

Page 3: ...Preventing System BEPS detects when the batter ies level becomes weak When the batteries are getting low a clear indication is displayed on LCD to warn the user of this condition However the meter co...

Page 4: ...efore turning the meter off When the meter is OFF detach the electrode from the meter before immersing it in the storage solution SPECIFICATIONS Range pH 4 00 to 19 99 pH mV 600 0 mV 2000 mV autorangi...

Page 5: ...When the CFM symbol blinks the read ing is stable and calibration can be confirmed 8 Press CFM to confirm the first buffer 9 If everything is satisfactory the LCD will display Stor and then the expect...

Page 6: ...e above procedure only when readings are stable 7 To exit the calibration mode and return to normal measurement press TXCAL mV CALIBRATION Contact your nearest Hanna Service Center for periodic mV cal...

Page 7: ...nterface cable connected between HI 504920 and HI 504910 step 1 it can happen that the HI 504910 must be located in a place difficult to reach when the HI 504920 pH probe is immersed in the solution I...

Page 8: ...0 Calibration alarm time out 01 to 99 days OFF OFF 20 Auto Off Power down time out ON OFF ON 21 Firmware version 22 Battery level test 90 HI 504910 terminal mode 99 Password2 0000 to 9999 0000 Setup c...

Page 9: ...y checks for entered time date accuracy as follows 0 hh 23 0 mm 59 01 dd 28 29 30 31 1 MM 12 1998 YYYY 2097 2 To change the password the correct code must be entered first If the password has been for...

Page 10: ...When turned on the meter checks if the time out time has expired If yes the message Cal date scrolls across the LCD and the DATE tag blinks as a reminder LAST CALIBRATION DATA Last calibration data ar...

Page 11: ...ers In this case check the correct polarity of the adapter before connecting it to the meter BATTERY REPLACEMENT The LCD will then display the last pH calibration date Press DISPLAY to scan remaining...

Page 12: ...4 00 12 50 13 00 13 50 14 00 0 10 0 14 0 20 0 10 0 18 0 29 0 40 Note To prevent damage to the electrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach...

Page 13: ...torage Solution Follow the Preparation Procedure above before taking measurements Note NEVER STORE THE ELECTRODE IN DISTILLED WATER OR DRY PERIODIC MAINTENANCE Inspect the electrode and the cable The...

Page 14: ...SOLUTIONS HI 7091L Reducing Pretreatment solution 500 mL bottle HI 7092L Oxidizing Pretreatment solution 500 mL bottle ELECTRODES HI 62920 pH electrode with built in temperature sensor matching pin Ti...

Page 15: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Reviews: