ACCESSORIES
pH CALIBRATION SOLUTIONS
HI 70004P
pH 4.01 buffer, 20 mL sachet (25 pcs)
HI 70007P
pH 7.01 buffer, 20 mL sachet (25 pcs)
HI 70010P
pH 10.01 buffer, 20 mL sachet (25 pcs)
HI 7004L
pH 4.01 buffer, 500 mL bottle
HI 7006L
pH 6.86 buffer, 500 mL bottle
HI 7007L
pH 7.01 buffer, 500 mL bottle
HI 7009L
pH 9.18 buffer, 500 mL bottle
HI 7010L
pH 10.01 buffer, 500 mL bottle
ELECTRODE STORAGE SOLUTION
HI 70300L
Storage solution, 500 mL bottle
ELECTRODE CLEANING SOLUTIONS
HI 70000P
Rinsing solution, 20 mL sachet (25 pcs)
HI 7061L
General cleaning solution, 500 mL bottle
HI 7073L
Protein cleaning solution, 500 mL bottle
HI 7074L
Inorganic cleaning solution, 500 mL bottle
HI 7077L
Oil & Fat cleaning solution, 500 mL bottle
REFILL ELECTROLYTE SOLUTIONS
HI 7071
3.5M KCl + AgCl Electrolyte, 50 mL (4 pcs)
HI 7072
1M KNO
3
Electrolyte, 50 mL (4 pcs)
HI 7082
3.5M KCl Electrolyte, for double junction electrodes, 50
mL (4 pcs)
• Dry Membrane/Junction:
Soak in
Storage Solution HI 70300
for at least 1 hour.
• Drifting:
Soak the electrode tip in warm Hanna Solution
HI
7082
for 1 hour, then flush tip with distilled water. Refill with
fresh
HI 7071
for single junction electrodes and
HI 7082
for
double junction electrodes.
• Low Slope:
Refer to the cleaning procedure above.
• No Slope:
Check the electrode for cracks in glass stem or bulb
and replace the electrode.
• Slow Response/Excessive Drift:
Soak the tip in Hanna Solu-
tion
HI 7061
for 30 minutes, rinse thoroughly in distilled water
and then follow the Cleaning Procedure above.
Replace the solution in the protective cap with a few drops of
HI
70300 Storage Solution
. Follow the Preparation Procedure above
before taking measurements.
Note:
NEVER STORE THE ELECTRODE IN DISTILLED WATER OR DRY.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to
the meter must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb.
Connectors must be perfectly clean and dry.
If any scratches or cracks are present, replace the electrode.
Rinse off any salt deposits with water.
For refillable electrodes:
Refill it with fresh electrolyte (
HI 7071
for single junction or
HI 7082
for double junction electrodes). Allow the electrode to stand upright
for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
General Soak in Hanna
HI7061
General Cleaning Solution
for
approximately 1 hour.
Removal of films, dirt or deposits on the membrane/junction:
-
Protein
Soak in Hanna
HI 7073
Protein Cleaning
Solution
for 15 minutes.
-
Inorganic
Soak in Hanna
HI 7074 Inorganic Cleaning
Solution
for 15 minutes.
-
Oil/grease
Rinse with Hanna
HI 7077 Oil and Fat Clean
ing Solution
.
IMPORTANT:
After performing any of the cleaning procedures rinse
the electrode thoroughly with distilled water, drain and refill the
reference chamber with fresh electrolyte, (not necessary for GEL filled
electrodes) and soak the electrode in
HI 70300
Storage Solution
for
at least 1 hour before taking measurements.
TROUBLESHOOTING
Evaluate your electrode performance based on the following possibilities:
• Noise
(Readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction:
Refer to the Cleaning Procedure above.
- Loss of shielding
due to low electrolyte level (in refillable
electrodes only): refill with fresh
HI 7071
for single junction or
HI
7082
for double junction electrodes.