24
RELATIVE mV CALIBRATION (HI 111
AND HI 113)
Press
CAL
when the instrument is in RELATIVE mV measurement mode.
The mV and tags will blink. Absolute mV is displayed on the
primary LCD and AbS message is displayed on the secondary LCD.
When the absolute reading is stable and in measurement range, the
instrument asks for confirmation.
If the reading is out of range, WRONG tag will be displayed.
Press
CFM
to confirm the absolute value. The instrument will display 0.0
mV on the primary LCD and rEL message on the secondary LCD. In this
moment the relative mV offset is equal to absolute mV reading.
Use the
ARROW
keys if you want to change the displayed relative mV value.
Press
CFM
to confirm the relative mV value. The relative mV offset is
displayed on the primary LCD and OFF message on the secondary LCD for
a few seconds. The instrument returns to measurement mode.
Note:
The relative mV value can be changed only inside the relative mV offset
window (± 2000 mV).