Stand-Alone Tester (Up to Three-Point Calibration)

1.  Rinse the electrode tip with purified water and blot dry. Then rinse with the buffer being used for calibration.

2.  Press and hold ON/OFF button until “

CAL

 

MODE

” is displayed.

3.  When “

7.01 

USE

” or “

6.86 

USE

” is displayed with “CAL” tag blinking, place the tip of the electrode in the 

correct buffer.

CAL

MODE

   

7.01

USE

4.  When the buffer has been recognized, “

REC

“ is displayed. 

WAIT

” is displayed with the stability indicator (

#

) blinking until the reading is stable. Wait until the measurement is 

stored and the stability indicator disappears.

 

7.01

REC

   

#

7..01

WAIT

5.  To save a one-point calibration and return to measurement mode, press the ON/OFF button.

CAL

 

SAVE

” is displayed briefly.

With 2P option selected

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

7.  After the second point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to measurement 

mode automatically.

With 3P option selected

 

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

7.  Place the tip of the electrode in the third buffer rinse beaker, then in the third calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

8.  After the third point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to measurement 

mode automatically.

With Hanna Lab App (Up to Four-Point Calibration)

Connect the tester to the Hanna Lab App and follow the calibration procedure.

See the App Help section for the calibration procedure.

Summary of Contents for HALO 2 HI9810362

Page 1: ...INSTRUCTION MANUAL HI9810362 Wireless pH Tester for Meat with built in specialized electrode Foodcare ...

Page 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our contact list at www hannainst com Dear Customer...

Page 3: ...H Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8 10 Care Maintenance 9 Refilling the Electrode 9 Storage 9 11 Warning Error Messages 9 12 Abbreviations 9 13 Acces...

Page 4: ...0 0 F Resolution pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers Hanna pH 1 68 4 01 7 01 10 01 or NIST pH 1 68 4 01 6 86 9 18 Temperature compensation Automatic ATC or Manual MTC Electrode Body material Polyvinylidene Fluoride PVDF Glass Low Temperat...

Page 5: ...ite ultraviolet light and fungal growth the PVDF body also has high abrasion resistance and mechanical strength The conical tip allows for easy penetration into solids and semisolids The PVDF outer junction sleeve can be removed and cleaned Once cleaned a small amount of supplied gel electrolyte is added and the junction is refreshed improving the measurement and extending the life of the tester T...

Page 6: ...ed battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup button to enter Setup mode and navigate menu items To exit Setup mode press the Setup button after SEt TIME option is displayed Press the ON OFF button to configure menu item ...

Page 7: ...hen the tester is in discoverable mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measurement Once connected to the application the ON OFF button can be pressed to tag the current reading Press the ON OFF button from measurement mode The display shows SEt TAG fo...

Page 8: ...de in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE is displayed briefly and the tester will return to measurement mode automatically With 3P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibra...

Page 9: ...ridge electrolyte B Replace the sleeve Ensure the O ring is fixed inside the electrode Rinse off excess gel with distilled or deionized water Shake the probe down gently to eliminate any trapped air bubbles Soak the probe in HI70300 Electrode storage solution for a minimum of 30 minutes before calibrating Storage To ensure a quick response the glass tip and junction should be kept hydrated When no...

Page 10: ...his product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in hazardous environments WARRANTY HI9810362 is warranted for a period of one year against defects in workmanship and materi...

Page 11: ...ovide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interferenc...

Page 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810362 Printed in ROMANIA ...

Reviews: