To view the EC calibration data, press GLP when the instrument is in EC
measurement mode. The instrument will display the calibration standard and
the temperature of the calibrated standard. Use the ARROW keys to scroll
through the calibration data displayed on the 3rd LCD line.
The cell factor in cm
-1
determined from the
calibration with the current reading.
The calibration offset factor in μS/cm together with
the current reading.
The Solution Standard and calibration temperature.
The temperature coefficient set at the calibration
moment together with the current reading.
The reference temperature together with the
current reading.
EC/TDS
GLP
Information