124
PHOSPHATE LOW RANGE
PHOSPHATE LOW RANGE
PHOSPHATE LOW RANGE
PHOSPHATE LOW RANGE
PHOSPHATE LOW RANGE
Phosphate LR
SPECIFICATIONS
Range
0.00 to 2.50 mg/L
Resolution
0.01 mg/L
Accuracy
±0.04 mg/L ±4% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 610 nm
Method
Adaptation of the Ascorbic Acid method. The
reaction between phosphate and the reagent
causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93713-0
Powder reagent
1 packet
REAGENT SETS
HI 93713-01
Reagents for 100 tests
HI 93713-03
Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Phosphate LR on
the secondary LCD by pressing PROGRAM
W
and
V
.
• Rinse, cap and shake the cuvet several times
with unreacted sample. Fill the cuvet with
10 mL of sample up to the mark and replace
the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
10 mL