6. Troubleshooting
6.2 Error messages
Error 7
Software
* Contact a Service Engineer of your local Hanil Scientific’s partner and
get the firmware upgrade. Wire disconnection or tuning of the
instrument must be performed only by a Service Engineer authorized
by Hanil Scientific’s partner.
Error 8
Imbalance
* Check the balance status of the samples in the rotor and turn off
and on the instrument to check the status.
* If the error message does not disappear, please contact a Service
Engineer of your local Hanil Scientific’s partner.
Error 9
RPM Sensor
Engineer of your local Hanil Scientific’s partner.
sales
@ihanil.com / [email protected] l 18
Error
Action
Possible Causes
* If the installed software have any bugs, this message may come up.
* If the rotor recognition fails, this message comes up.
* The message will be cleared by coupling an appropriate rotor
* Disassemble and couple a compatible rotor and restart
the instrument to check out the status.
* If the error message does not disappear, please contact a Service