S E R V I C E M A N U A L
System
RoMedic
TM
Troubleshooting
The green light blinks, but stops after a while.
Automatic calibration complete
The green light blinks.
- Manual calibration is required, see section on manual
calibration.
- Check to ensure that no load was applied to the
handlebar when the lift was started e.g., that nothing was
hanging on the handlebar. In that case, remove it and the
green light will stop blinking.
All or parts of the battery scale blink quickly.
Something is wrong with the lift’s electrical system.
Restart EvaDrive. If there is still a problem, contact
Handicare.
Red light + battery scale blink
- Restart
- Contact Handicare
If unusual sounds are heard:
- Try to locate the source of the sound.
- Take the lift unit out of service and contact Handicare.
The lift cannot be started
- Check the battery switch.
- Check battery connection.
- Connect charger and charge for at least 2 hours.
- Contact Handicare.
Resetting service indicator
•
The handle should be ON.
•
Press and hold the start button untill the service indicator starts flashing (aprox 2 sec. )
•
Without releasing the start button, press down both of the manouver buttons for the rear wheels and hold
untill the service indicator is turned off (aprox. 5 sec).