Changing of spare parts
80100172 Handle EvaDrive
Set battery switch OFF. Disconnect battery cable. Remove battery. Detach cover. The cover is secured by pads of velcro.
Remove Cable entry. Separate cable entry. Detach connectors from control boxes (VR2 box AND ADM-box) Remove
cable from cover.
Note
: routing of cables. Remove 2 screws from handle bracket.
Replace Handle including hinge assembly. Replace 2x Screws (K6S M8x16). Adjust handle height. Tighten screws.
Thread cable through cover and routing them to control boxes. Attach connectors to VR2 and ADM-box (see p. 9).
Assemble cable entry and adjust cable positions and length.
Note
: Cable length adjustment is important in relation to
mast and handle position. Fit cable entry in cover. Reposition the cover and press it into place. Reattach battery and
battery cable. Important! If a new handle is is fitted, an initial calibration must be performed as follows:
-
Make sure the handle is not physically loaded in any way (e.g. handset must not be hangin on the handle)
-
Start the system by pressing the ON button
-
The Snail light starts blinking with double beats
-
Allow the system to initiate until the Snail light stops blinking. This may take up to 10 minutes
-
Test all basic functions of the drive system. Make sure approximately the same force is needed to initiate
forwards and backwards motion. If not, perform a manual calibration of the system (se page 10 in the user
manual).
60900022 Battery Eva Drive
Set battery switch OFF. Detach battery cable. Slide the old battery upwards. If necessary adjust handle position.
Replace with new battery. Gently slide battery back into place. Reattach battery cable and switch ON. Readjust handle
position.
80100174 Cover leg EvaDrive
Replace cover leg by pressing it into back of leg profile, use LocTite715 glue to fasten the profile.
80100175/176 Cover Eva450Drive/ Cover Eva600Drive
Set battery switch OFF. Disconnect battery cable. Detach cover. The cover is secured by pads of velcro. Remove the
velcro from base. Cut the new velcro into eight pieces. Attach the bits of velcro to the same place on the base as the old
ones. Keep them together with their ”twins”. Detach Cable entry. Remove the emergency stop button by unscrewing
the plastic retainer on the inside of the cover.
Note
routing of cables. Remove and replace cover.
Route cables and fit cable entry into cover. Re-connect cables to control boxes, 4x to VR2 and 1x to ADM-box (see p.
9). Remove the glue cover from the velcro. Secure the cover by pressing it into place. Re attach battery cable and switch
ON.
80100196 Emergency stop button EvaDrive
Set battery switch to OFF. Disconnect battery cable. Detach cover. The cover is secured by pads of velcro. Detach Cable
entry. Remove the emergency stop button by unscrewing the plastic retainer on the inside of the cover. Note routing of
cables. Detach emergency stop cable connector from control box VR2. Disconnect cables from emergency stop button
by unscrewing the cable clamp screws.
NOTE!
numbers of connection points (no 21, 22). Remove and replace emer-
gency stop button with the new one, re-attach Emergency stop cables to the emergency stop button. Route cables and
fit cable entry into cover. Re-connect cables to control boxes, 4x to VR2 and 1x to ADM-box (see p. 9). Secure the cover
by pressing it into place. Re attach battery cable and switch ON.