background image

12

Subject to change without notice

8.4 Test of the connection to the oscilloscope

For the test of the connection of the host (PC) to the Ethernet 
interface of the oscilloscope, please go into the menu “Start“ and 
select “Run“. Start the command interpreter by the input of the 
instruction “cmd“ into the input fi eld. Complete the sequence 
by the “Enter“ key or confi rm the input by clicking OK.

 

It opens an input window. After the input character you give to 
confi rm the instruction “ping 192.168.10.10“ (in the represented 
example the Ethernet interface of the oscilloscope has the IP 
address 192.168.10.10) and confi rm the input with Enter.

 

If the Ethernet interface answers the example represented by 
the oscilloscope to the “Ping“ instruction without errors as 
in, the connection is correct. If the interface does not answer 
attainable e.g. with an error message, no connection is present 
or the connection is disturbed. In this case please examine all 
network cables between oscilloscope and host (PC), as well 
as the preset interface parameters of the Ethernet interface 
with the oscilloscope and the Ethernet LAN interface with the 
host (PC). If the connection over further network elements e.g. 
switches, routers, network servers, etc. is used, examine if 
necessary these further connections, as well as the presets of 
the appropriate network elements.

9. Application

In combination with HMLab 1.0 or higher, the interface HO730 
can be used via Ethernet as well as USB. The HMLab settings 
have to correspond with the interface settings of the oscil-
loscope. With HMLab an oscilloscope or several oscilloscopes 
can be operated at the same time, if these are connected by a 
common LAN network. Likewise an oscilloscope or also several 
oscilloscopes can be operated at the same time, if they are atta-
ched over USB or the LAN-/WAN networks are distributed over 
several LAN-/WAN networks among themselves with switches 
and/or routers (e.g. over the Internet) connected.

The interface HO730 has also a Web server, which can be used 
with a Web browser (e.g. Internet Explorer, …). The following func-
tions are supported by the Web server (only in digital mode):
–  Indicate the equipment data
–  Selections of the Readout and save over Web browser func-

tion

–  Control and inquiry with programming instructions (SCPI 

instructions) over input mask

Indicate the equipment data

 

Selections of the Readout and save over Web browser func-
tion

 

E t h e r n e t  c o n f i g u r a t i o n          

Summary of Contents for HO730

Page 1: ...E T H E R N E T U S B I N T E R F A C E H O 7 3 0 Installation guidelines English...

Page 2: ...nditions in individual cases HAMEG Instruments GmbH KONFORMIT TSERKL RUNG DECLARATION OF CONFORMITY DECLARATION DE CONFORMITE DECLARACI N DE CONFORMIDAD Hersteller Manufacturer Fabricant Fabricante HA...

Page 3: ...ruction 5 5 1 Removing the existing interface 5 5 2 Fitting the interface HO730 5 6 Interface Selection 6 7 USB Driver Installation 7 7 1 Installation under Windows XP 7 8 Ethernet con guration 9 8 1...

Page 4: ...e is equipped with a Type A USB female connector For direct connection with a host controller or an indirect con nection via a USB hub a USB cable is required equipped with Type B male connector one e...

Page 5: ...of the oscilloscope to equalise potentials between oscilloscope and your body Maintain this connection during the tting removing Only touch the interface at its mounting panel 5 1 Removing the existi...

Page 6: ...activated With the activating of Ethernet the menu option Parameter also ap pears additional in the Readout which the Ethernet interface parameter settings made possible see section 8 Reference The ho...

Page 7: ...splayed Only in this case the USB Driver must be installed The USB Driver is located in the directory HO730_D2xx_Dri ver 7 1 Installation under Windows XP 7 1 1 Please choose No not this time and clic...

Page 8: ...7 1 3 4 After the selected path is displayed click Next 7 1 4 Thereafter the Hardware Installation window is displa yed showing a warning to continue the installation procedure As this warning is irr...

Page 9: ...s is used as network IP address and the last IP address is used as Broadcast IP address in each case two IP addresses have to be taken off from the number of possible host addresses see table 1 Privat...

Page 10: ...on control to the desired value The menu option Save in the Readout saves the preset interface parameters by the oscilloscope and or determined automatically after press the menu option Save with acti...

Page 11: ...he input eld Subnet mask please enter from your or from a network administrator spe ci ed subnet mask for the PC e g 255 255 255 0 In the input eld Default gateway please enter your or from a network...

Page 12: ...the host PC If the connection over further network elements e g switches routers network servers etc is used examine if necessary these further connections as well as the presets of the appropriate n...

Page 13: ...e found in the Internet www hameg com for downloading and updating your PC The path of the software is Products Oscilloscopes oscilloscope type Software Firmware below the oscilloscope picture Softwar...

Page 14: ...14 Subject to change without notice...

Page 15: ...15 nderungen vorbehalten A p p l i c a t i o n...

Page 16: ...riestra e 6 A Rohde Schwarz Company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de authorized...

Reviews: