S p e c i f i c a t i o n s
HM8123E/151007/ce · Subject to alterations · © HAMEG Instruments GmbH · ® Registered Trademark · DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49 (0) 6182 800 100 · www.hameg.com · info
@
hameg.com
A Rohde & Schwarz Company
Accessories supplied:
Power cable, Operator’s Manual, HZ184 4 Terminal
Kelvin Test Cable and HZ188 4 Terminal SMD Component Test Fixture
Optional accessories:
HZ181 4 Terminal Test Fixture including Shorting Plate
HZ186 4 Terminal Transformer Test Cable
HO880 IEEE-488 (GPIB) Interface
Measurement conditions and functions
Test frequency:
20 Hz to 200 kHz (69 steps)
Frequency accuracy:
±100 ppm
AC test signal level:
50 mV
rms
–1.5 V
rms
Resolution:
10 mV
rms
Drive level accuracy:
± (5 % + 5 mV)
Internal Bias Voltage:
0 to +5.00 V
dc
Resolution:
10 mV
External Bias Voltage:
0 to + 40 V
dc
(fused 0.5 A)
Internal Bias Current:
0 to +200 mA
Resolution:
1 mA
Ranging:
Auto and Hold
Trigger:
Continuous, manual or external via inter-
face, Handler Interface or Trigger Input
Trigger delay time:
0 to 999 ms in 1 ms steps
Measurement time (f
≥
1 kHz)
FAST
70 ms
MEDIUM
125 ms
SLOW
0.7 s
Others Instruments Functions
Test signal level monitor:
Voltage, current
Error Correction:
Open, Short, Load
Save / Recall:
9 instrument settings
Front-end Protection:
V
max
‹
√
2/C
@
V
max
‹ 200 V, C in Farads
(1 Joule of stored energy)
Low Potential and
Ground, Driven Guard or Auto (fused)
Low Current Guarding:
Constant Voltage Mode (25 Ω source)
Temperature effects:
R, L or C:
± 5ppm/°C
Interface:
USB/RS-232, optional IEEE-488
Safety Class:
Safety Class I (EN61010-1)
Power supply:
110–230 V ± 10 % / 50–60 Hz
Power consumption:
approx. 20 Watt
Operating temperature:
+ 10 °C to + 40 °C
Max. rel. humidity:
10%–90 % (without condensation)
Dimensions (W x H x D):
285 x 75 x 365 mm
Weight:
approx. 4 kg
LCR-Bridge HM8118
Valid at 23 °C after a 30 minute warm-up period
Conditions
Test signal voltage:
1 V
rms
Open and short corrections performed
Measurement time:
SLOW
Display
Measurement modes:
Auto, L+Q, L+R, C+D, C+R, R+Q, Z+
θ
, Y+
θ
,
R+X, G+B, N-
θ
, M
Equivalent circuits:
Auto, Series or Parallel
Parameters displayed:
Value, Deviation or % Deviation
Averaging:
2 - 99 measurements
Accuracy
Primary Parameter:
Basic accuracy (Test voltage: 1.0V,
measurement SLOW/MEDIUM, autoranging
mode, constant voltage OFF, bias off)
For FAST mode double the basic accuracy
values
Impedance:
Secondary Parameter:
Basic accuracy DQ
± 0.0001
@
f = 1 kHz
Phase angle
± 0.005°
@
f= 1 kHz
Ranges
|Z|, R, X:
0.01 mΩ to 100 MΩ
|Y|, G, B:
10 nS to 1000 S
C:
0.01 pF to 100 mF
L:
10 nH to 100 kH
D:
0.0001 to 9.9999
Q:
0,1 to 9999.9
θ
:
-180° to +180°
Δ
:
-999.99 % to 999.99 %
M:
1 μH to 100 H
N:
0.95 to 500
0.2 % + |Z|/1.5 GΩ
0.05 % +
|Z|/2 GΩ
0.1 % +
|Z|/1,5 GΩ
0.1 % + 1 mΩ/|Z|
0.3 % + 1 mΩ/|Z|
0.5 % +
|Z|/100 MΩ
0.2 % +
|Z|/100 MΩ
0.2 % +
2 mΩ/|Z|
0.5 % +
2 mΩ/|Z|
0.5 % +
5 mΩ/|Z|
+
|Z|/10MΩ
10 mΩ
20 Hz
1 kHz
10 kHz
100 kHz
2,5 Ω
100 Ω
25 kΩ
1 MΩ
4 MΩ
100 MΩ