Hameg HM8118 User Manual Download Page 29

65

Measuring Equipment

be tested by connecting a simple, passive component (e.g. 
1kOhm resistance) to the instrument. In the BIN menu, set 
a generous pass/fail criterion and measure the voltage of 
PIN 25 (BIN2 on the circuit, BIN1 in the instrument) for the 
25-pole plug to PIN 1 (GND). In case of „Pass“, the voltage
should be 0V, in the case of „Fail“, it should correspond to
the external voltage which must be applied to PIN 9.

For detailed information on the binning interface in the 
context of the PIN and jumper assignment, refer to the 
HO118 manual at www.hameg.com.

8.5.2  HO118 Description

The Store/Recall feature enables you to determine a ma-

ximum of 9 binning configurations. Binning configura

-

tions can also be operated via remote control interface. 
The bridge HM8118 can sort components in up to 8 sepa-
rate bins: six pass sorting bins, one secondary parameter 
sorting bin and one general sorting bin for errors. At any 
given time, only one sorting bin (BIN) is activated.

Output signal:

Negative TRUE, open collector, opto-isolated, selectable 
pull-ups.

Measurement types:

Since the HM8118 is used for classification, the number of 

BIN Type

Description

0...5

Pass BIN

This sorting bin is used if the measured value is 

within the user-defined bin limit. If the measured 

value is within this limit, it will be assigned to bin 0 

(BIN 0). Outside the limit that is defined for bin 0, 

the assignment within the limit for bin 1 (BIN 1) is 
performed. This process is repeated until the limit 
for bin 5 (BIN 5) is exceeded. If the measured value 

exceeds the defined range limits for bins 1 to 5, it 

will be assigned to the General-Failure bin.

6

Secondary 
Parameter 
Failure BIN

This sorting bin is used if the primary value is 
within the range for the sorting bins 0 ... 5 and 
only the secondary parameters exceed the limit for 
sorting bin 6.

7

General Fai-
lure BIN

This sorting bin output is activated if the sorting 

does not apply to one of the first 7 bins.

measurement types is limited to the modes needed for the 
characterization of components.

R-Q: Resistance value and quality

C-D: Capacitance value and loss angle

L-Q: Inductance and quality

Sorting bins (BINs):

Pass bin: Bin 0...5 for primary parameters

Fail bin: Bin 6 for secondary parameters, bin 7 for general
errors (General Failure BIN).

Maximum current for an output voltage of 1 V is 15 mA.

Index:

Analog measurement completed.

Measurement completed:

Complete measurement completed.

Alarm:

Notification about a known bug.

External trigger: 

Opto-isolated, selectable pull-up, pulse width >10µs.

8.5.3   Sorting Bin Preferences (BINs)

The HM8118 must be in manual mode. Select the respec-
tive function of the parameter to be sorted. As mentioned 
in the section „Measurement Types“, all functions can be 
used. To be able to enter binning parameters, press the 
MENU key and select the option BIN. A binning interfact 
must be integrated to be able to access the binning menu.

Example: 

Binning: 

 ON

BIN Number: 

0

BIN: 

 Open

Nominal:

100.0

Low limit: 

-4.0%

High limit: 

+5.0%

Binning ON/OFF:

ON: Binning function activated

OFF: Binning function deactivated

BIN Number:

Selecting the BIN number

Bins 0 to 5 correspond to the primary pass bins

Bin 6 corresponds to the secondary parameter failure bin

Bin 7 (General Failure BIN 7) does not have a menu entry.

BIN OPEN or CLOSED:

OPEN: The respective BIN is activated.

CLOSED: The respective BIN is deactivated.

❙ At least the first bin must be activated.

Nominal value of the classification:

Enter the nominal value with the number keys and

confirm with the Enter key.

The new value and the associated units will be displayed.
A nominal value for bin 6 is not applicable.

Fig. 8.16: HO118 internal circuitry

Summary of Contents for HM8118

Page 1: ...HM8118 Programmable LCR Bridge User Manual...

Page 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Page 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Page 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Page 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Page 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Page 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Page 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Page 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Page 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Page 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Page 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Page 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Page 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Page 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Page 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Page 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Page 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Page 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Page 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Page 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Page 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Page 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Page 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Page 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Page 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Page 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Page 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Page 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Page 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Page 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Page 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Page 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Page 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Page 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Page 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Reviews: