Measuring Equipment

Due to its design, the measurement adapter HZ188 has a 
fringing capacitance, a residual inductance and a residual 
resistance which impact the accuracy of the measured va-
lues. To minimize these impacts, the compensation of im-
pedance errors caused by adapters becomes necessary.

To compensate or eliminate this measurement error, it is 
recommended to perform an open and short compensa-
tion (OPEN/SHORT compensation) with the LCR bridge 
HM8118. For the open compensation, for the test adapter 
HZ188 you must loosen the screw on the right side coun-
terclockwise and then push the right contact pin to the 
right until both contact pins are electrically open. The re-
sulting gap between the contact pins must correspond 
to the dimensions of the SMD component to be mea-
sured. Fixate the right contact pin by turning the screw 
clockwise.

For the short compensation, for the test adapter HZ188 
you must loosen the screw on the right side counterclock-
wise and then push the right contact pin to the left un-
til both contact pins are electrically connected. Fixate the 
right contact pin by turning the screw counterclockwise.

For frequency dependent components, make sure to perform 
an OPEN and SHORT compensation for each of the 69 test 
frequencies.

10

Änderungen vorbehalten

Für  den  „Kurzschlussabgleich“  ist  bei  dem  Messadapter 

HZ188 die Schraube auf der rechten Seite gegen den Uhr-

zeigersinn zu lösen und anschließend der rechte Kontaktstift 

mit der Taste nach links zu drücken, bis beide Kontaktstifte 

elektrisch  verbunden  sind.  Danach  ist  der  rechte  Kontakt-

stift durch drehen der Schraube im Urzeigersinn zu fixieren 

(siehe Bild 2).

Bild 2 „Kurzschlußabgleich“

Screw

Lever

Without SMD 
Component

Fig. 8.13: Open compensation with HZ188

The compensation values that are measured during the 
compensation process will be stored in the memory of the 
LCR bridge HM8118 and are valid until another compensa-
tion is performed. If any changes to the measurement se-
tup are implemented, it becomes necessary to perform 
a new compensation. For more information about the 
OPEN/SHORT compensation, see chapter 7.2.

8.5  Sorting Components with Option HO118 

Binning Interface

A binning interface (25 pol. interface) is particularly useful 
for a production environment:

To test incoming components, e.g. for incoming goods,

To select components by limit,

To test multiple components with similar values

The HO118 binning interface enables the use with exter-
nal hardware which sorts components by physical type af-
ter the HM8118 measurement. Data lines for eight sorting 
containers and control lines are intended (ALARM, INDEX, 
EOM,TRIG).

8.5.1  HO118 Circuit

The HM8118 with integrated HO118 binning interface is al-
ways delivered in a condition that allows for an external 

power supply to be connected. Specifically, this means 

that jumper J1 is on position 2-3, jumper J3 on position 1-2 
and the DIP switch set to „OFF“. These settings deactivate 
the internal pull-ups.

The following conditions must be be met to operate the 
binning interface:

Use external pull-ups.

Provide external power supply between 5V and 40V.

The circuit is „active low“, i.e.the voltage drops to 0V as 
soon as the criterion for the respective BIN (set in the ins-
trument) is met. The function of the binning interface can 

The test adapter HZ188 allows SMD components of up to a size 
of 0603 to 1812 (in inches) to be tested. This corresponds to a 
size of approximately 1.6mm to 4.5mm.

Fig. 8.15: Optional equipment HO118 (binning interface)

We recommended the factory-installed option HO118. Otherwise,
it becomes necessary to open the instrument which would break
the warranty seal which in turn would void the warranty.

10

Änderungen vorbehalten

Für  den  „Kurzschlussabgleich“  ist  bei  dem  Messadapter 

HZ188 die Schraube auf der rechten Seite gegen den Uhr-

zeigersinn zu lösen und anschließend der rechte Kontaktstift 

mit der Taste nach links zu drücken, bis beide Kontaktstifte 

elektrisch  verbunden  sind.  Danach  ist  der  rechte  Kontakt-

stift durch drehen der Schraube im Urzeigersinn zu fixieren 

(siehe Bild 2).

Bild 2 „Kurzschlußabgleich“

Screw

Locked

Lever

Fig. 8.14: Short compensation with HZ188

Summary of Contents for HM8118

Page 1: ...HM8118 Programmable LCR Bridge User Manual...

Page 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Page 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Page 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Page 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Page 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Page 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Page 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Page 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Page 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Page 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Page 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Page 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Page 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Page 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Page 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Page 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Page 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Page 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Page 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Page 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Page 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Page 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Page 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Page 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Page 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Page 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Page 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Page 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Page 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Page 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Page 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Page 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Page 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Page 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Page 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Reviews: