Hameg HM8118 User Manual Download Page 26

62

Measuring Equipment

The measurement adapter HZ186 is designed for measure-
ments of transformers or transmitters in combination with 
transformer measurement functions of the LCR bridge 
HM8118. The measurement adapter is directly connected 
to the front panel BNC connectors of the LCR bridge via 
the four BNC sockets. 

The 4-wire transformer test lead is a convenient tool to 
measure the mutual inductance (M), the transformer ra-
tio (N) and the phase difference 

Θ

 in a frequency range 

between 20Hz and 200kHz of a transformer or transmit-
ter. The measurement adapter serves as interface between 
the LCR bridge and the four included test leads. For the 
measurement, the transformer / transmitter to be measu-
red is connected to the measurement adapter via test lead, 
according to the imprinted wiring on the primary and the 
secondary side. This equipment is optional and is not in-
cluded in delivery.

8.3.1  Compensation HZ186

Due to their design, the test lead HZ186 and the connected 
test leads have a fringing capacitance, self inductance and 
self-resistance which impacts the accuracy of the measu-

1

3

2

2

1  Transformer test adapter
2  Test lead for high number of windings
3  Test lead for low number of windings

Fig. 8.7: 4-wire transformer test lead

In case of a faulty measurement, the LCR bridge does not display 
any value for N.

red values. To minimize these impacts, the compensation 
of impedance errors caused by adapters and leads beco-
mes necessary.

To compensate or eliminate this measurement error, it is 
recommended to perform an open and short compensa-
tion (OPEN/SHORT compensation) with the LCR bridge 
HM8118. For the open compensation, the four test leads 
are connected to the measurement adapter HZ186. Be-
fore starting the open compensation, the two black test 
leads (which are connected to the „COMMON“ BNC con-
nectors) are connected. It is also necessary to connect the 
two red test leads that are connected to the BNC connec-
tors „N“ and „1“. 

For the short compensation, the two red test leads and the 
two black test leads are connected to each other.

8.3.2  Transformer Measurement

The measurement of a transformer can always result in va-
rying measurement results. This is related both to the iron 
core losses as well as to the unknown state of the prema-
gnetized core. The component to be measured is depen-
dent on the frequency as well as on the applied measure-
ment voltage. The measuring instrument determines the 
values for L, R and C by measuring the impedance and the 
related phase angle. The angle determines an inductive, 
capacitive or real value (L,C,R). Consequently, the amount 
of the impedance increases as the voltage increases, and 
the phase angle is heavily dependent on the measurement 
frequency (due to change in magnetization and iron core 
loss and visible in the „Z-Theta“ mode [7]). If a transformer 
is measured as „open“, the measurement values are plau-
sible. However, if the secondary side is short-circuited, it is 
only possible to measure considerably fewer measurement 
values. The values in case of a short-circuited secondary 
page correspond nearly precisely to the core losses.

8.3.3  Mutual Inductance

To measure the mutual inductance, the HM8118 applies 
the same procedure as for the regular inductance. Instead 
of measuring the voltage via primary winding, the voltage 
will be measured at the secondary transformer winding.

For frequency dependent components, make sure to perform 
an OPEN and SHORT compensation for each of the 69 test 
frequencies.

11

Änderungen vorbehalten

Für den „Kurzschlussabgleich“ sind die beiden roten Mess-

leitungen und die beiden schwarzen Messleitungen gemein-

sam miteinander zu verbinden.

Bild 2:  OPEN-Abgleich = trennen

SHORT-Abgleich = kurzschließen

Bild 3 „Primär- und Sekundär Transformator“

Bild 4 „Spar-Transformator“

HAMEG HZ186

COMMON

N

1

trennen

kurzschließen

kurzschließen

HAMEG HZ186

COMMON

N

1

Transformator

rot

schwarz

HAMEG HZ186

COMMON

N

1

Spar-

Transformator

Gemeinsamer

Masseanschluss

Fig. 8.8: OPEN / SHORT calibration with HZ186

Short Circuit

Short Circuit

Open

HAMEG HZ186

Technical Data HZ186

Function: 

Measurement adapter to operate (via 4-wire 
connection) with LCR bridge HM8118

Measurable components: Transformers, transmitters
Measurable parameters: 

Mutual inductance M (1 µH...100 H), 
Transformer ratio N (0,95...500), 

phase difference φ between primary and se

-

condary winding (-180° to +180°)

Frequency range:

20 Hz  to  200 kHz

Test lead length:

approximately  35 cm

Connectors:

BNC sockets (4), BNC connectors (4)

Safety standards:

EN61010-1; IEC61010-1; 
EN61010-031; IEC61010-031

Environmental conditions: Contamination Class 2, internal use
Operating temperature:

+5° C to +40 °C

Storage temperature:

-20 °C to +70 °C

Weight:

approximately  240 g

Summary of Contents for HM8118

Page 1: ...HM8118 Programmable LCR Bridge User Manual...

Page 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Page 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Page 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Page 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Page 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Page 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Page 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Page 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Page 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Page 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Page 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Page 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Page 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Page 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Page 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Page 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Page 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Page 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Page 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Page 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Page 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Page 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Page 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Page 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Page 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Page 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Page 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Page 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Page 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Page 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Page 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Page 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Page 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Page 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Page 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Reviews: