background image

Subject to change without notice

13

T

3

: alters the middle frequencies

T

4

: alters the leading edge

T

5

: alters the lower frequencies

(LF)

(LF)

One output provides 

0.2V

pp

 ±1% (t

r

 <4ns) for 10:1 probes,

and the other 

2V

pp

 ±1%  for 100:1 probes. When the

attenuator     switches are set to 

5mV/div

 vertical deflection

coefficient, these calibration voltages correspond to a
screen amplitude of 

4div

.

The output sockets have an internal diameter of 4.9mm to
accommodate the internationally accepted shielding tube
diameter of modern Modular Probes and F-series slimline
probes. Only this type of construction ensures the extremly
short ground connections which are essential for an
undistorted waveform reproduction of non-sinusoidal high
frequency signals.

Adjustment at 1kHz

The C-trimmer adjustment compensates the capacitive
loading on the oscilloscope input (approx. 20 pF for the
HM 303). By this adjustment, the capacitive division
assumes the same ratio as the ohmic voltage divider to
ensure the same division ratio for high and low frequencies,
as for DC. (For 1:1 probes or switchable probes set to 1:1,
this adjustment is neither required nor possible). A baseline
exactly parallel to the horizontal graticule lines is a major
condition for accurate probe adjustments. (See also “Trace
rotation 

TR

”).

Connect the probes (Types HZ51, 52, 53, 54, or HZ36) to
the 

CH.I

 input. All pushbuttons should be released (in the

out position). Set input coupling to DC, the attenuator to

5 mV/div.

, and 

TIME/DIV.

 switch to 

0.2 ms/div.

, and all

variable controls to 

CAL.

 position. Plug the the probe tip

into the appropriate calibrator output socket, i.e. 10:1
probes into the 

0.2V

 socket, 100:1 probes into the 

2V

socket.

1 kHz

incorrect correct incorrect

Approximately 2 complete waveform periods are displayed
on the CRT screen. Now the compensation trimmer has to
be adjusted. Normally, this trimmer is located in the probe
head. On the 100:1 probe HZ53, however, it is located in the
connecting box at the other end of the cable. Adjust the
trimmer with the insulating screw driver provided until the
tops of the square wave signal are exactly parallel to the
horizontal graticule lines (see 1 kHz diagram). The signal
height should then be 4 div. ± 0.12div. (= 3 %). During this
adjustment, the signal edges will remain invisible.

Adjustment at 1MHz

Probes HZ51, 52 and 54 can also be HF-compensated.
They incorporate resonance de-emphasing networks (R-
trimmer in conjunction with inductances and capacitors)
which permit probe compensation in the range of the
upper frequency limit of the vertical oscilloscope amplifier.

HZ51, HZ54

Only this compensative adjustment ensures optimum
utilisation of the full bandwidth, together with constant
group delay at the high frequency end, thereby reducing
characteristic transient distortion near the leading edge
(e.g. overshoot, rounding, ringing, holes or bumps) to an
absolute minimum.
Using the probes HZ51, 52 and 54, the full bandwidth of
the HM303 can be utilized without risk of unwanted
waveform distortion.
Prerequisite for this HF compensation is a square wave
generator with fast risetime (typically 4 ns), and low
output impedance (approx. 50

), providing 0.2V and 2V at

a frequency of approx. 1MHz. The calibrator output of the
HM303 meets these requirements when the 

CAL.

pushbutton is depressed.
Connect the probe to CH.I input. Depress the 

CAL.

pushbutton for 1MHz. All other pushbuttons should be
released (out position). Set the CH.I input coupling to 

DC

,

attenuator switch to 

5mV/div

, and TIME/DIV. switch to

0.2µs/div.

 Set all variable controls to 

CAL

. position.

Insert the probe tip into the output socket marked 

0.2V

.

A waveform will be displayed on the CRT screen, with
leading and trailing edges clearly visible. For the HF-
adjustment now to be performed, it will be necessary to
observe the rising edge as well as the upper left corner of
the pulse top. The connecting boxes of the HZ51 and
HZ54 contain one R-trimmer screw each, while that of the
HZ52 provides three. These R-trimmers have to be adjusted
such that the beginning of the pulse is as straight as
possible. Overshoot or excessive rounding are unaccept-
able. This is relatively easy on the HZ51 and HZ54, but
slightly more difficult on the HZ52. The rising edge should
be as steep as possible, with a pulse top remaining as
straight and horizontal as possible.
On the HZ52, each of the three trimmers has a clearly
defined area of influence on the waveform shape (see
Fig.), offering the added advantage of being able to
straighten out waveform abberations near the leading
edge.

Summary of Contents for HM 303-4

Page 1: ...ge 8 Amplitude Measurements 8 Time Measurements 9 Connection of Test Signal 10 First Time Operation 12 Trace Rotation TR 12 Probe compensation and use 12 Operating Modes of the Y Amplifier 14 X Y Oper...

Page 2: ...ual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection to a cable Basica...

Page 3: ...WG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied...

Page 4: ...nal Features Component Tester 1kHz 1MHz Calibrator OSCILLOSCOPES Specifications Vertical Deflection Operating modes Channel I or II separate both Channels alternated or chopped Chopper frequency appro...

Page 5: ...prox 500M DC 1MHz max 15kV DC peakAC HZ47 Viewing Hood for Oscilloscopes HM205 408 604 1 2 1005 and 1007 HZ48 Viewing Hood for Oscilloscopes 303 304 305 604 3 and 1004 during transpor tation of an osc...

Page 6: ...is standard It has left the factory in a safe condition This instruction manual contains important information and warnings which have to be followed by the user to ensure safe operation and to retain...

Page 7: ...tment designation isstatedforpossiblequeries thishelpstowards speeding up the processing of guarantee claims Maintenance Variousimportantpropertiesoftheoscilloscopeshouldbe carefully checked at certai...

Page 8: ...However forsignalmagnitudesandvoltage designations in oscilloscope measurements the peak to peak voltage Vpp value is applied The latter corresponds to the real potential difference between the most...

Page 9: ...he attenuator series resistor will break down causingdamagetotheinputoftheoscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal...

Page 10: ...must be divided by 10 The fall time of a pulse can also be measured by using this method The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement tr...

Page 11: ...obecompensation page M7 Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time In all cases where the oscilloscope bandwidth must be fully utilized e g...

Page 12: ...ion required The HM303 accepts all signals from DC direct voltage up to a frequency of at least 30MHz 3dB For sinewave voltages the upper frequency limit will be 50MHz 6dB However in this higher frequ...

Page 13: ...should then be 4 div 0 12div 3 During this adjustment the signal edges will remain invisible Adjustment at 1MHz Probes HZ51 52 and 54 can also be HF compensated They incorporate resonance de emphasing...

Page 14: ...the 3 pushbuttons CH I II DUAL and ADD in the Y field of the front panel For Mono mode all 3 buttons mustbeintheirreleasedpositions onlychannelIcanthen beoperated ThebuttonCHI II TRIG I IImustbedepres...

Page 15: ...in the X Y mode can exceed an angle of 3 above 120 kHz It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the...

Page 16: ...o buttons Y CH I 20mV div AC TIME DIV 0 2ms div Triggering NORMAL with LEVEL setting internal or external triggering Figure 2 Amplitude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the...

Page 17: ...ing mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the LEVEL control Its adjusting range is directly dependent on the display...

Page 18: ...eringcanalsobeusefultodisplaysignalsbelowthe triggerthreshold lessthan0 5div Itisthereforeparticularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in...

Page 19: ...iodes marked OVERSCAN which are located between the attenuators Should one LED illuminate without an input signal this means that the respective vertical positioning control has been improperly adjust...

Page 20: ...orizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approxi mately evaluated The determination of actual values will come with...

Page 21: ...n by 180 degrees round about the center point of the scope graticule In Circuit Tests Caution During in circuit tests make sure the circuit is dead No power from mains line or battery and no signal in...

Page 22: ...ort circuit Resistor 510 Junction B C Junction B E Mains transformer prim Capacitor33 F Junction E C FET Z diode below 7V Z diode beyond 7V Diode paralleled by 680 2 Diodes antiparallel Silicon diode...

Page 23: ...n in the edge zone of the screen must be accepted It is limited by standards of the CRT manufacturer The same is valid for tolerances of the orthogonality the undeflected spot position the non lineari...

Page 24: ...anneldisplaywiththe CHI IIbutton is unnecessary it is contained indirectly in the tests above stated input of the vertical amplifier e g using a x1 probe the displayed signal in the50mV div position v...

Page 25: ...splay should not shift horizontally during a change of the trigger coupling from AC to DC with a sine wave signal without DC offset If both vertical inputs are AC coupled to the same signal and both t...

Page 26: ...LDOFF knob Component Tester After pressing the COMP TESTER button a horizontal straight line has to appear immediately when theCOMP TESTER socket is open The length of this trace should be approx 8div...

Page 27: ...d from all voltage sources Normally the capacitors are discharged approx 6 seconds after switching off However with a defective instrument an interruption oftheloadisnotimpossible Therefore afterswitc...

Page 28: ...g Checktoseethatallcircuitboardconnections are making good contact and are not shorting to an adjacent circuit Especially inspect the connections between the PCBs to front chassis parts to CRT PCB to...

Page 29: ...ring TRIG EXT button depressed sync signal 0 3Vpp to 3Vpp to TRIG INP socket Line triggering TRIG selector switch to Select trigger coupling with TRIG selector switch Trigger frequency ranges AC 20Hz...

Page 30: ...nector horizontal deflection in X Y mode Y POS II Controls vertical position knob of channel II display Inoperative in X Y mode POWER Turns scope on and off pushbutton LED LED indicates operating cond...

Page 31: ...1 3 5 6 7 8 9 14 15 16 17 2 4 10 11 12 13 20 21 22 23 24 25 26 27 28 30 31 34 36 37 38 33 32 18 29 19 35...

Reviews: