background image

12

Subject to change without notice

First Time Operation

Before applying power

 to the oscilloscope it is recom-

mended that the following simple procedures are
performed:

• Check that all pushbuttons are in the 

out

 position, i.e.

released.

• Rotate the variable controls with arrows, i.e. 

TIME/DIV.

variable control, 

CH.I

 and 

CH.II

 attenuator variable

controls, and 

HOLD OFF

 control to their calibrated

detent.

• Set all controls with marker lines to their midrange

position (marker lines pointing vertically).

• The 

TRIG.

 selector lever switch in the X-field should be

set to the position uppermost.

• Both 

GD

 input coupling pushbutton switches for 

CH.I

and 

CH.II

 in the Y-field should be set to the 

GD

 position.

Switch on the oscilloscope by depressing the red 

POWER

pushbutton. An LED will illuminate to indicate working
order. The trace, displaying one baseline, should be visible
after a short warm-up period of approx. 10 seconds.
Adjust 

Y-POS.I

 and 

X-POS.

 controls to center the baseline.

Adjust 

INTENS.

 (intensity) and 

FOCUS

 controls for

medium brightness and optimum sharpness of the trace.
The oscilloscope is now ready for use.

If only a spot appears (

CAUTION!

 CRT phosphor can be

damaged), reduce the intensity immediately and check
that the 

XY

 pushbutton is in the released (out) position. If

the trace is not visible, check the correct positions of all
knobs and switches (particularly 

AT/NORM.

 button in out

position).

To obtain the maximum life from the cathode-ray tube, the
minimum intensity setting necessary for the measurement
in hand and the ambient light conditions should be used.

Particular care is required when a single spot is
displayed,

 as a very high intensity setting may cause

damage to the fluorescent screen of the CRT. Switching
the oscilloscope off and on at short intervals stresses the
cathode of the CRT and should therefore be avoided.

The instrument is so designed that even incorrect operation
will not cause serious damage. The pushbuttons control
only minor functions, and it is recommended that before
commencement of operation all pushbuttons are in the
“out” position. After this the pushbuttons can be operated
depending upon the mode of operation required.

The HM303 accepts all signals from DC (direct voltage) up
to a frequency of at least 30MHz (

3dB). For sinewave

voltages the upper frequency limit will be 50MHz (

6dB).

However, in this higher frequency range the vertical
display height on the screen is limited to approx. 4-5div.
The time resolution poses no problem. For example, with
50MHz and the fastest adjustable sweep rate (10ns/div.),
one cycle will be displayed every 2div. The tolerance on
indicated values amounts to ±3% in both deflection
directions. All values to be measured can therefore be
determined relatively accurately.

However, from approximately 10MHz upwards the
measuring error will increase as a result of loss of gain. At
18MHz this reduction is about 10%. Thus, approximately
11% should be added to the measured voltage at this
frequency. As the bandwidth of the amplifiers may differ
slightly (normally between 30 and 35MHz), the measured
values in the upper frequency range cannot be defined
exactly. Additionally, as already mentioned, for frequencies
above 30MHz the dynamic range of the display height
steadily decreases. The vertical amplifier is designed so
that the transmission performance is not affected by its
own overshoot.

Trace Rotation TR

In spite of Mumetal-shielding of the CRT, effects of
the earths magnetic field on the horizontal trace
position cannot be completely avoided. This is
dependent upon the orientation of the oscilloscope
on the place of work. A centred trace may not align
exactly with the horizontal center line of the graticule.
A few degrees of misalignment can be corrected by a
potentiometer accessible through an opening on the
front panel marked TR.

Probe compensation and use

To display an undistorted waveform on an oscilloscope,
the probe must be matched to the individual input
impedance of the vertical amplifier.

For this purpose a square wave signal with a very fast rise
time and minimum overshoot should be used, as the
sinusoidal contents cover a wide frequency range. The
frequency accuracy and the pulse duty factor are not of
such importance.

The built-in calibration generator provides a square wave
signal with a very fast risetime (<4ns), and switch-
selectable frequencies of approx. 1kHz and 1MHz from
two output sockets below the CRT screen.

This signal should not be used for frequency cali-
bration!

Summary of Contents for HM 303-4

Page 1: ...ge 8 Amplitude Measurements 8 Time Measurements 9 Connection of Test Signal 10 First Time Operation 12 Trace Rotation TR 12 Probe compensation and use 12 Operating Modes of the Y Amplifier 14 X Y Oper...

Page 2: ...ual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection to a cable Basica...

Page 3: ...WG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied...

Page 4: ...nal Features Component Tester 1kHz 1MHz Calibrator OSCILLOSCOPES Specifications Vertical Deflection Operating modes Channel I or II separate both Channels alternated or chopped Chopper frequency appro...

Page 5: ...prox 500M DC 1MHz max 15kV DC peakAC HZ47 Viewing Hood for Oscilloscopes HM205 408 604 1 2 1005 and 1007 HZ48 Viewing Hood for Oscilloscopes 303 304 305 604 3 and 1004 during transpor tation of an osc...

Page 6: ...is standard It has left the factory in a safe condition This instruction manual contains important information and warnings which have to be followed by the user to ensure safe operation and to retain...

Page 7: ...tment designation isstatedforpossiblequeries thishelpstowards speeding up the processing of guarantee claims Maintenance Variousimportantpropertiesoftheoscilloscopeshouldbe carefully checked at certai...

Page 8: ...However forsignalmagnitudesandvoltage designations in oscilloscope measurements the peak to peak voltage Vpp value is applied The latter corresponds to the real potential difference between the most...

Page 9: ...he attenuator series resistor will break down causingdamagetotheinputoftheoscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal...

Page 10: ...must be divided by 10 The fall time of a pulse can also be measured by using this method The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement tr...

Page 11: ...obecompensation page M7 Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time In all cases where the oscilloscope bandwidth must be fully utilized e g...

Page 12: ...ion required The HM303 accepts all signals from DC direct voltage up to a frequency of at least 30MHz 3dB For sinewave voltages the upper frequency limit will be 50MHz 6dB However in this higher frequ...

Page 13: ...should then be 4 div 0 12div 3 During this adjustment the signal edges will remain invisible Adjustment at 1MHz Probes HZ51 52 and 54 can also be HF compensated They incorporate resonance de emphasing...

Page 14: ...the 3 pushbuttons CH I II DUAL and ADD in the Y field of the front panel For Mono mode all 3 buttons mustbeintheirreleasedpositions onlychannelIcanthen beoperated ThebuttonCHI II TRIG I IImustbedepres...

Page 15: ...in the X Y mode can exceed an angle of 3 above 120 kHz It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the...

Page 16: ...o buttons Y CH I 20mV div AC TIME DIV 0 2ms div Triggering NORMAL with LEVEL setting internal or external triggering Figure 2 Amplitude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the...

Page 17: ...ing mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the LEVEL control Its adjusting range is directly dependent on the display...

Page 18: ...eringcanalsobeusefultodisplaysignalsbelowthe triggerthreshold lessthan0 5div Itisthereforeparticularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in...

Page 19: ...iodes marked OVERSCAN which are located between the attenuators Should one LED illuminate without an input signal this means that the respective vertical positioning control has been improperly adjust...

Page 20: ...orizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approxi mately evaluated The determination of actual values will come with...

Page 21: ...n by 180 degrees round about the center point of the scope graticule In Circuit Tests Caution During in circuit tests make sure the circuit is dead No power from mains line or battery and no signal in...

Page 22: ...ort circuit Resistor 510 Junction B C Junction B E Mains transformer prim Capacitor33 F Junction E C FET Z diode below 7V Z diode beyond 7V Diode paralleled by 680 2 Diodes antiparallel Silicon diode...

Page 23: ...n in the edge zone of the screen must be accepted It is limited by standards of the CRT manufacturer The same is valid for tolerances of the orthogonality the undeflected spot position the non lineari...

Page 24: ...anneldisplaywiththe CHI IIbutton is unnecessary it is contained indirectly in the tests above stated input of the vertical amplifier e g using a x1 probe the displayed signal in the50mV div position v...

Page 25: ...splay should not shift horizontally during a change of the trigger coupling from AC to DC with a sine wave signal without DC offset If both vertical inputs are AC coupled to the same signal and both t...

Page 26: ...LDOFF knob Component Tester After pressing the COMP TESTER button a horizontal straight line has to appear immediately when theCOMP TESTER socket is open The length of this trace should be approx 8div...

Page 27: ...d from all voltage sources Normally the capacitors are discharged approx 6 seconds after switching off However with a defective instrument an interruption oftheloadisnotimpossible Therefore afterswitc...

Page 28: ...g Checktoseethatallcircuitboardconnections are making good contact and are not shorting to an adjacent circuit Especially inspect the connections between the PCBs to front chassis parts to CRT PCB to...

Page 29: ...ring TRIG EXT button depressed sync signal 0 3Vpp to 3Vpp to TRIG INP socket Line triggering TRIG selector switch to Select trigger coupling with TRIG selector switch Trigger frequency ranges AC 20Hz...

Page 30: ...nector horizontal deflection in X Y mode Y POS II Controls vertical position knob of channel II display Inoperative in X Y mode POWER Turns scope on and off pushbutton LED LED indicates operating cond...

Page 31: ...1 3 5 6 7 8 9 14 15 16 17 2 4 10 11 12 13 20 21 22 23 24 25 26 27 28 30 31 34 36 37 38 33 32 18 29 19 35...

Reviews: