If there is a component failure, do not attempt to open the Controller or Lamp enclosure of the EQ-
400.  

Any service to the system must be performed only by factory authorized and trained technicians.  To avoid 
injury, under no circumstances should the user open or modify the Lamp or Controller enclosure. 

7. 

Troubleshooting Guide 

If any errors are detected, a front panel indicator LED will be lit, indicating the source of the fault. The below 
table describes the fault conditions and possible remedies. 

Indicator 

Fault condition 

Possible remedies 

LASER 

OVERTEMP 

Laser internal temperature too 

high 

1.

 

Check for sufficient Controller cooling water flow 

2.

 

Ensure that cooling water temperature is between 18 
and 24 degC 

LASER FAULT 

Laser module internal fault 

1.

 

Cycle power to the PCU 

2.

 

Contact Energetiq if condition persists 

PCU FAULT 

Controller internal temperature 
too high 

1.

 

Confirm that cooling fan is operating 

2.

 

Ensure that fan opening and front panel vents are not 
obstructed 

PCU FAULT 

Laser power not reaching setpoint 

1.

 

Contact Energetiq 

PCU FAULT 

Internal power supply voltage low 

1.

 

Contact Energetiq 

PCU FAULT 

Lamp failed to ignite 

1.

 

If lamp was previously running, allow to cool for 5 
minutes then try again 

2.

 

Contact Energetiq if condition persists 

LAMPHOUSE 
OVERTEMP 

Lamphouse internal temperature 
too high 

1.

 

Check for sufficient lamphouse cooling water flow 

2.

 

Ensure that cooling water temperature is between 18 

and 24 degC 

LAMPHOUSE 
INTERLOCK 

Control Cable not connected 
properly 

1.

 

Ensure that the cable is plugged in and retaining 
screws tightened at both ends (Lamp and Controller) 

LAMPHOUSE 
INTERLOCK 

Laser fiber not connected 
properly 

1.

 

Ensure that the collimator bracket is seated in the 
lamphouse, and the screws are properly tightened 

LAMPHOUSE 

INTERLOCK 

Lamphouse cover loose or 

missing 

1.

 

Tighten or replace cover 

2.

 

NEVER attempt to operate the unit with the lamphouse 
cover removed 

EXTERNAL 

INTERLOCK 

External interlock open 

1.

 

Ensure that pins 5 and 13 of the remote I/O connector 

are jumpered, or that pin 13 re5VDC from an 
external supply 

2.

 

Check any external interlock switches that may be 

connected 

COVER 
INTERLOCK 

Controller cover(s) not in place 

1.

 

Ensure that covers are in place and screws properly 
tightened 

2.

 

NEVER attempt to operate the unit with the Controller 
covers removed 

 

Additionally, if the bulb fails to ignite within 6 seconds after the OPERATE switch was activated, the 
LASER ON LED will go out, the PCU FAULT LED will be lit, and LAMP ON LED will remain off.  If this 
occurs, turn the OPERATE switch to the OFF position (down) and begin at Step 1 of start up again. If this 
occurs multiple times, contact Energetiq service. 

Summary of Contents for ENERGETIQ LDLS EQ-400

Page 1: ...Model EQ 400 Laser Driven Light Source LDLS Operation Manual Revision 9 May 2022 Part Number DOC 7388...

Page 2: ...ical information including drawings schematics and specifications contained in this document are the property of Energetiq and shall not be reproduced in whole or in part without the written consent o...

Page 3: ......

Page 4: ...Remote Interface 10 3 6 External interlock connection 10 4 EQ 400 Setup 10 4 1 Connections 10 4 2 Chiller Information 11 5 EQ 400 Operation 12 5 1 Startup 12 5 2 Stopping 12 5 3 Manual Reset 12 6 Serv...

Page 5: ...Section 4 for connection details For additional information contact Energetiq Technology Technical Support Services at 1 781 939 0763 X111 or e mail EUVproductsupport Energetiq com 2 Safety instructi...

Page 6: ...not open the enclosure of either the Lamp or the Controller Dangerous invisible infrared laser beams and hazardous voltages exist inside the Lamp Opening the chassis both voids the warranty and expose...

Page 7: ...on 400W 1070nm pump laser under a failure mode condition where aperture is itself accessible 4 meters Required Optical Density OD for safety eyewear 6 or higher at 1070 nm These values are based on a...

Page 8: ...and Safety Notification The following safety labels appear on the product Table 1 shows the location of each label on the EQ 400 system and Table 1 describes their meanings Figure 1 EQ 400 Safety lab...

Page 9: ...e outlet aperture for laser radiation 4 UV Hazard warning label indicates hazardous levels of UV light are present 5 Protective housing label warns of an enclosed laser radiation hazard 6 Explanatory...

Page 10: ...00 The shipping box should contain EQ 400 Power Supply Controller unit EQ 400 Lamp unit Accessory Box Interconnecting cable from Lamp to Power Supply Controller 9 pin D sub AC Power cable AC Power plu...

Page 11: ...dicators are described below in Table LED Label Meaning when lit POWER OK AC power is connected to the EQ 400 PCU LAMP ON UV Light is on LASER ON Laser power is ON and laser light is being produced wi...

Page 12: ...unctions This connector also functions as the remote interlock connector 4 RS 485 Connector 9 pin D sub Connector for optional RS 485 interface See Sections 3 and 4 for electrical details and commands...

Page 13: ...mp 3 Control Connector 9 pin D sub Provides various power and control signals to from the Power Supply Controller No other connector or cable may be used with the EQ 400 other than the one supplied 4...

Page 14: ...he Controller weighs 18 8 kg 41 5 lb A two person lift is recommended Use care when moving the Controller not to bend the fiber optic cable sharply Minimum bend radius is 50mm 2 Connect the Lamp unit...

Page 15: ...rlock switch to the interlock pins of the remote I O connector functions as remote interlock connector If using remote control mode connect user s control system to the remote I O connector using a su...

Page 16: ...conds the laser will switch to full power 5 2 Stopping The following steps explain the proper procedures stopping To turn the lamp off simply turn the OPERATE Switch to the OFF position It is recommen...

Page 17: ...rgetiq if condition persists LAMPHOUSE OVERTEMP Lamphouse internal temperature too high 1 Check for sufficient lamphouse cooling water flow 2 Ensure that cooling water temperature is between 18 and 24...

Page 18: ...resentative SMC PN HRS018 A 10 T PN Description EQ 400 CHILLER KIT EQ 400 Chiller with Tubing Kit Note Cooling fluid is not included and must be locally sourced 9 Warranty Energetiq Technology Inc s w...

Page 19: ...15 EQ 400 Controller top in mm 10 1 2 Lamp Dimensional Drawings EQ 400 Side view in mm...

Page 20: ...y non condensing 80 max for temperatures up to 31 C decreasing linearly to 50 max at 35 C Pollution Degree 2 normally only non conductive pollution occasional temporary condensation possible Installat...

Page 21: ...n when ON LAMP MODULE FAULT 3 Pulled to digital common when OK float on FAULT CONTROLLER FAULT 4 Pulled to digital common when OK float on FAULT ISOLATED 5V SUPPLY 5 200mA maximum referenced to digita...

Page 22: ...face Commands The RS 485 serial interface is provided for user adjustment of laser operating power to increase or decrease light output The EQ 400 is shipped with the laser power set to a factory defa...

Page 23: ...actory default value Power XXX X B Saves present settings to flash memory This command should be issued after the settings are at their desired values If not changes will be lost if power to the EQ 40...

Page 24: ...1 out of range 0 OK 5 Laser current source 1 fault 0 OK 6 Lamphead temperature 1 overtemperature 0 OK 7 PCU temperature 1 overtemperature 0 OK 8 Lamphead interlock 1 open 0 OK 9 External interlock 1...

Page 25: ...21 Energetiq Technology Inc 205 Lowell St Wilmington MA 01887 Phone 1 781 939 0763 Fax 1 781 939 0769 Email info energetiq com www energetiq com DOC 7388 Rev 9 5 22...

Reviews: