KXX-A12040 A

 

Take  precautions  to  protect  the  window  from  contamination  or  scratches  when  packing  or  shipping 
equipment in which the product is installed. 
 
 

(2) Precautions when mounting 

 

The  product  is  a  precision  optical  component,  so  placing  excessive  loads  on  the  package  may  cause 
problems  such  as  package  warping  or  damage,  chip  peeling,  wire  breakage,  damage  to  thermoelectric 
cooler,  or  poor  optical  characteristics.  Take  full  precautions  when  mounting  the  product  on  the  board, 
inserting the product into the socket, or installing the heatsink. 

 

Never handle the product in the following ways since this may cause the package to break. 

  apply repeated stress to the leads. 

  suddenly cool or suddenly heat the product. 

 

Check  the  pin  connection  positions  when  mounting  the  product  in  the  socket.  Installing  the  product 
reversely or in the wrong position may damage the product. Be sure to turn on the power after the product 
has been properly installed. 

 

Be  aware  that  inserting  the  product  reversely,  inserting  it  incorrectly,  or  causing  short-circuit  between 
terminals may lead to malfunctions. 

 

If external force is repeatedly applied to the leads, this may damage the leads. When installing this product 
in locations subject to vibration, secure it with resin or a holder, etc. 
 

(3) Precautions for designing the circuit board 

 

The packaging of the micro-spectrometer is electrically conductive, so be careful when designing the circuit 
to avoid short circuit caused by contact with a circuit pattern. 
 

Soldering 
(1) Precautions during soldering 

 

Take sufficient care to make sure that the soldering iron tip temperature and soldering time are correct. Do 
not attempt soldering at high temperatures or for long periods. Soldering at high temperatures or for long 
periods may cause the package to crack or the window material to separate from the package.   

 

It  is  recommended  that  the  MS  series  be  soldered  within  3  seconds  at  370  °C  or  less  and  the 
micro-spectrometers within 3.5 seconds at 350 °C or less when soldering by hand, and both be soldered 
within 10 seconds at 260 °C or less when using reflow soldering.   

 

Take  measures  to  prevent  solder  or  flux  from  flying  outward  and  sticking  to  the

 

light  input  window, 

contaminating it. 

 
(2) Flux 

 

Use non-cleaning solder or rosin type flux. Using flux with relatively strong acid or alkali levels or inorganic 
flux may cause corrosion on the leads. 

 
(3) When using a soldering iron 

 

Set the soldering iron tip temperature by referring to the recommended soldering condition. If you cannot 
provide these conditions, then grip the root of the lead you are soldering with tweezers or a similar tool to 
prevent heat from conducting to the product package. 

 

Summary of Contents for C12666MA

Page 1: ...when no light is input the product generates an output consisting of an offset component and dark signal The offset component is not dependent on integration time but the dark signal is dependent on...

Page 2: ...e supplied with it Use caution when turning on the power supply because an inrush current flows at that time Table 1 shows the maximum current in steady state and typical inrush current Table 1 Maximu...

Page 3: ...h a wrist strap having protective resistance of 750 k to 35 M directly to the skin and ground the strap If the wrist strap does not include protective resistance there is a risk of electric shock haza...

Page 4: ...ductive case Avoid using plastic or styrofoam as they may generate static electricity by vibration during shipping etc causing breakdown or deterioration of the heed type Use a conductive carrying cas...

Page 5: ...fluctuations so separate it as far as possible from the video signal line and voltage supply line Apply the input clock pulses at the specific timing and voltage to the product connection terminals on...

Page 6: ...the optical fiber SMA connector and then use the slotted holes to secure the mini spectrometer at a position where the optical fiber is free from stress Figure 3 Precautions when connecting input opt...

Page 7: ...cure it with resin or a holder etc 3 Precautions for designing the circuit board The packaging of the micro spectrometer is electrically conductive so be careful when designing the circuit to avoid sh...

Page 8: ...d of time seal it in a moisture proof package containing silica gel or keep it in a low humidity desiccator 4 Input optical fibers option When connecting the input optical fiber clean the ferrule on t...

Reviews: