The electrical terminals of the head type are connected to the internal image sensor signal terminals. Take the 
following precautions on static electricity management, circuit board design, and handling, etc. 
 

3-1. Static electricity management 

 
Electrostatic sensitive products come with an electrostatic warning label on the product packing. Handling of 
the product requires taking precautions on the following points to avoid product damage and deterioration due 
to static electricity. 
 

Figure 1

  Electrostatic warning label 

 

 

 
(1) Workplace and facilities, etc. 

 

Lay a conductive mat (750 k

 to 1 G

) on the surface of the workbench and ground it. 

 

Use conductive flooring material or lay a conductive mat on the workplace floor and ground it. 

 

Ground all manufacturing equipment and inspection devices. 

 

Keep moisture at approximately 50%. Low humidity tends to cause static electricity and high humidity is 
prone to moisture absorption. 

 
(2) Handling 

 

Using an ionizer or similar item to eliminate electrical charges is recommended when handling the product. 

 

Wear anti-static clothing and conductive shoes (100 k

 to 100 M

). 

 

Attach a wrist strap (having protective resistance of 750 k

 to 35 M

Ω

) directly to the skin, and ground the 

strap. If the wrist strap does not include protective resistance, there is a risk of electric shock hazard due to 
electric leak. Also wear conductive finger sacks or gloves. 

 

Tools such as tweezers used to handle the product may sometimes become electrically charged. Connect a 
ground line as needed. 

 

Use  a  soldering  iron  with  an  insulation  resistance  of  10  M

Ω

  or  higher.  The  soldering  iron  tip  should  be 

grounded. 

 

If the product is induction-charged and contacts with a metal, excessive current may flow due to electrostatic 
discharge causing damage to the product. To prevent induction charging, keep objects (insulators such as 
plastic and vinyl, PC display monitors and keyboards, etc. that may possibly be electrically charged) away 
from the product. The product may be induction-charged even by just bringing such objects close to the 
product.  If  keeping  such  objects  near  the  product  is  unavoidable,  then  use  an  ionizer,  etc.  to  remove 
electrostatic charges from the objects that are apt to be electrostatically charged. 

 

Friction  on  the  product  causes  electrostatic  charges.  If  such  friction  is  unavoidable,  then  remove  the 
electrostatic charges using an ionizer, etc. 

ATTENTION

OBSERVE PRECAUTIONS

FOR HANDLING

ELECTROSTATIC

SENSITIVE

DEVICES

Summary of Contents for C12666MA

Page 1: ...when no light is input the product generates an output consisting of an offset component and dark signal The offset component is not dependent on integration time but the dark signal is dependent on...

Page 2: ...e supplied with it Use caution when turning on the power supply because an inrush current flows at that time Table 1 shows the maximum current in steady state and typical inrush current Table 1 Maximu...

Page 3: ...h a wrist strap having protective resistance of 750 k to 35 M directly to the skin and ground the strap If the wrist strap does not include protective resistance there is a risk of electric shock haza...

Page 4: ...ductive case Avoid using plastic or styrofoam as they may generate static electricity by vibration during shipping etc causing breakdown or deterioration of the heed type Use a conductive carrying cas...

Page 5: ...fluctuations so separate it as far as possible from the video signal line and voltage supply line Apply the input clock pulses at the specific timing and voltage to the product connection terminals on...

Page 6: ...the optical fiber SMA connector and then use the slotted holes to secure the mini spectrometer at a position where the optical fiber is free from stress Figure 3 Precautions when connecting input opt...

Page 7: ...cure it with resin or a holder etc 3 Precautions for designing the circuit board The packaging of the micro spectrometer is electrically conductive so be careful when designing the circuit to avoid sh...

Page 8: ...d of time seal it in a moisture proof package containing silica gel or keep it in a low humidity desiccator 4 Input optical fibers option When connecting the input optical fiber clean the ferrule on t...

Reviews: