[Value of frame rate (unit: frame/s)]
Binning
1×1 (Normal readout)
2×2 / 4×4
Horizontal
Vertical
1024 / 1536 /
2048
512
―
Free running mode
2048
30
100
100
1024
60
200
200
512
120
400
400
256
240
801
801
128
481
1603
1603
64
968
3206
3206
8
7894
25 655
25 655
External trigger mode
(Edge trigger)
2048
30
90
90
1024
60
164
164
512
120
278
278
256
240
427
427
128
481
583
583
64
712
712
712
8
884
884
884
External trigger mode
(Level trigger)
2048
30
90
90
1024
60
164
164
512
120
278
278
256
240
427
427
128
481
583
583
64
712
712
712
8
884
884
884
External trigger mode
(Synchronous trigger readout)
2048
30
98
98
1024
60
193
193
512
120
374
374
256
240
702
702
128
481
1251
1251
64
968
2052
2052
8
4664
4664
4664
Note
•
The calculation formula and the frame rate value of Start trigger mode (External
trigger mode) are same as the free running mode. About this mode, refer to 10-2-5-4
[Start trigger mode].