C11440-22CU/C11440-22CU01 Instruction manual_Ver.1.4
⑧
Power switch [POWER]
The power is turned on/off.
When the power switch is set to "ON", the camera turns on and starts initialization and the lamp
blinks in green.
When the initialization is completed , the lamp color stays in green.
When the camera transfers data and the lamp color is orange.
When the power switch is set to "OFF", the camera returns to the power off state and the lamp
turns off.
⑨
DC power input connector [DC IN]
This is the power supply terminal. Use the accessory AC adaptor.
⑩
USB 3.0 interface connector [USB 3.0]
This is connected to the USB 3.0 interface connector on the computer.
•
Do not connect Camera Link and USB 3.0 interface simultaneously.
•
When a connection interface is changed from Camera Link to USB 3.0, and vice
versa, the application software must be closed and the camera must be turned off.
⑪
Air outlet
This is the outlet for the heat ventilation.
•
To prevent overheating inside the camera, do not wrap the camera in cloth or
other material, or block the camera’s ventilation.
•
If the camera is being operated in an enclosed environment, ensure to keep
clearance at least 2 cm from both intake and exhaust vents when setting up.
14