C10910 INSTRUCTION MANUAL Ver.1.19E
71
.
11-3-8 M10913-01,-11+M10916-01
Vertical sweep
setting
All range
Horizontal sweep
setting
OFF
Gate mode
Cathode gate
MCP gate
Horizontal blanking
Operating
status
Max.
repetition
Operating status
Max.
repetition
Max. repetition
NORMAL
OPEN FIX
―
BACK SWP CUT
10 kHz *1
-
GATE
RISE TRIG.
10 kHz *1
RISE TRIG.
10 kHz *1
-
EXT.GATE
OPEN FIX
―
LEVEL TRIG.
10 kHz *2
-
Vertical sweep
setting
All range
Horizontal sweep
setting
Faster less than 100
µ
s
(100
µ
s is included.)
Gate mode
Cathode gate *3
MCP gate *4
Horizontal sweep
Operating
status
Max.
repetition
Operating status
Max.
repetition
Max. repetition
NORMAL
OPEN FIX
―
BACK SWP CUT
2 MHz
10 Hz
GATE
RISE TRIG.
10 kHz *1
RISE TRIG.
10 kHz *2
10 Hz
EXT.GATE
OPEN FIX
―
LEVEL TRIG.
10 kHz *2
10 Hz
Vertical sweep
setting
All range
Horizontal sweep
setting
Slower more than 200
µ
s
(200
µ
s is included.)
Gate mode
Cathode gate *3
MCP gate *4
Horizontal sweep
Operating
status
Max.
repetition
Operating status
Max.
repetition
Max. repetition
NORMAL
OPEN FIX
―
BACK SWP CUT
10 kHz *1
10 Hz
GATE
RISE TRIG.
10 kHz *1
RISE TRIG.
10 kHz *1
10 Hz
EXT.GATE
OPEN FIX
―
LEVEL TRIG.
10 kHz *2
10 Hz
Max. repetition of M10913-01,-11 is limited until 2 MHz, when horizontal sweep is enable.
It limits to when the cut of back sweep signal is unnecessary, and 4 MHz sweep also becomes possible. In that case,
please contact Hamamatsu subsidiary or local distributor
*1 When vertical frequency is below 10 kHz at the time of a GATE setup, if Gate trigger frequency surpasses 10 kHz,
a gate will be CLOSE FIX.
If vertical sweep frequency surpasses 10 kHz, it is not based on a trigger, but a gate will be
OPEN FIX.
*2 If a gate trigger surpasses 10 kHz, a MCP gate will be set to CLOSE FIX.
*3 Cathode gate open when vertical frequency is below 10 kHz, only while horizontal sweep is operated.
*4 MCP gate open, only while horizontal sweep is operated.