
P29 Instruction Manual
26
10.2 Replacing the fuse
You can replace the fuse to avoid unnecessary service and the associated device failure. The
fuse can be removed by hand. See figure 10-1 inside view P29
nicht gefunden werden.
.
The device must be disconnected from the power supply during the following steps.
To change the fuse, follow the steps below:
1. Make sure that the device is no longer connected to the power supply
2. Pull the fuse upwards.
3. Then reinsert the new fuse accordingly, only use the following fuse:
TR5 Fuse, Time-Lag type, Rated Current: 315mA, Voltage Rating: 250V
There is no further polarity to be considered here.
Servicing
The connecting hoses require monitoring and the seals must be inspected by the
operator at regular intervals to ensure they are still intact. Leaks can lead to
explosions.