P29 Instruction Manual
19
9 Technical data
Measurement data
Measurement range
See rating plate
Accuracy
0.5% +0.3 Pa of the measuring range
Resolution
Depends on the measuring range (max. 5 relevant
places)
Ambient conditions
Medium
Air, non-aggressive gases
Operating temperature
+10 °C to +50 °C
Storage temperature
-10 °C to +70 °C
Conformity
Declaration of conformity available upon request
Electrical data
Rated input
approx. 6 VA
Supply voltage
24VDC+/-10%
Output signal
0…+10 V (R
L
≥
2 k
Ω
) or-5…+5 V (R
L
≥
2 k
Ω
) or
0…20 mA (R
L
≤
500
Ω
) or4…20 mA (R
L
≤
500
Ω
)
Interface
RS-232
9600 Baud, 8 data bits, no parity, one stop bit
Appendix A
Parts in contact with measurement medium
Beryllium bronze CuBe2
Loctite 222
Mu metal (nickel alloy)
Vepuran Vu 4457/51
Brass CuZn39Pb3
UHU-Plus endfest 300 binder
Viton (tubing)
PA6.6 (valve)
NBR (valve)
PPS (valve)
Ferrite material M33 (Epcos)
Enamelled copper wire