P29 Instruction Manual
15
8.2.2.1 Value
This is used to set the volumetric flow value displayed at the maximum scaled pressure. 0 is
always used as the lower scale value on the volumetric flow display (root-extracted
measured value). The maximum pressure is the larger of the two scale values, which is used
as a basic calculation.
8.2.2.2 Unit
You can use this menu item to select the unit of the volumetric flow value. The following
volumetric flow units can be selected: m3/s, m3/h
8.2.3 Mass flow
This menu item is used to adjust the mass flow display values.
The following sub-menu items are available:
Value
Unit
8.2.3.1 Value
This is used to set the mass flow value displayed at the maximum scaled pressure. 0 is
always used as the lower scale value on the mass flow display (root-extracted measured
value). The maximum pressure is the larger of the two scale values, which is used as a basic
calculation.
8.2.3.2 Unit
You can use this menu item to select the mass flow unit. The following mass flow units can
be selected: kg/s, kg/min, kg/h
8.2.4 Flow rate
This menu item is used to adjust the flow rate display values.
The following sub-menu items are available:
Value
Unit
8.2.4.1 Value
This is used to set the flow rate value displayed at the maximum scaled pressure. 0 is always
used as the lower scale value on the flow rate display (root-extracted measured value). The
maximum pressure is the larger of the two scale values, which is used as a basic calculation.
8.2.4.2 Unit
Use this menu item to select the flow rate unit. The following flow rate units can be selected:
m/s, mph, f/s, f/min, km/h