4 Pneumatic connections
Connect the device to be calibrated to the KAL 84 using a Y connector as shown
below:
If the volume of the test object is high enough that the pressure bellows cannot
generate enough pressure, use an additional manual pump (included with the
calibrator upon request) and hook it up to the instrument with a Y connector.
Exercise caution when using the manual pump. A manual pump can quickly
generate high levels of pressure, which can destroy the instrument. Caution
is particularly advised when using instruments with low measurement
ranges.
5
Replacing the battery
The LCD will read “Low Battery” if available power falls below that required to operate
the instrument. When this happens, either replace the battery or recharge it using the
charger included with the calibrator. To replace the battery, open the cover of the
battery compartment located on the rear of the instrument.
Do not put defective batteries in the garbage. They must be taken to the
waste collection site specified by local authorities.