4  Pneumatic connections 

Connect the device to be calibrated to the KAL 84 using a Y connector as shown 

below: 

 

 

If the volume of the test object is high enough that the pressure bellows cannot 

generate enough pressure, use an additional manual pump (included with the 

calibrator upon request) and hook it up to the instrument with a Y connector. 

     Exercise caution when using the manual pump. A manual pump can quickly 

generate high levels of pressure, which can destroy the instrument. Caution 

is particularly advised when using instruments with low measurement 

ranges.

 

 

Replacing the battery 

The LCD will read “Low Battery” if available power falls below that required to operate 

the instrument. When this happens, either replace the battery or recharge it using the 

charger included with the calibrator. To replace the battery, open the cover of the 

battery compartment located on the rear of the instrument.

 

 

Do not put defective batteries in the garbage. They must be taken to the 

waste collection site specified by local authorities. 

 

Summary of Contents for KAL 84

Page 1: ...3554 12 2015 Instruction Manual KAL 84 Calibration Device halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup wal...

Page 2: ...6 2 2 User interfaces 6 2 2 1 The 4 digit LCD 7 2 2 2 ON OFF 7 2 2 3 Backlighting 7 2 2 4 Analog output 7 2 2 5 Units of pressure 7 2 2 6 Pressure ports 7 2 2 7 Purge valves 7 2 2 8 Automatic zero po...

Page 3: ...le throughout the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary materials...

Page 4: ...y technical personnel who are appropriately trained and authorized by the operator of the facility may assemble the instrument and set up its electrical connections The instrument may only be operated...

Page 5: ...Symbols The symbols given below are used throughout this manual to indicate instances when improper operation could result in the following hazards WARNING This warns you of a potential hazard that c...

Page 6: ...strument ideal for use in assembling pressure transducers as it allows the transducers to be calibrated on site Use of the KAL 84 for quality assurance purposes e g as a transfer standard for test equ...

Page 7: ...e This feature allows the operator to toggle between 2 units of pressure e g Pa and mmHg A red LED indicates the units used 2 2 6 Pressure ports These ports are for connecting the test object to the K...

Page 8: ...strument is run from the power adapter even if the KAL 84 is not turned on Because the battery may or may not be fully charged upon delivery we recommend charging the battery before start up The KAL 8...

Page 9: ...A manual pump can quickly generate high levels of pressure which can destroy the instrument Caution is particularly advised when using instruments with low measurement ranges 5 Replacing the battery...

Page 10: ...t least once a year A halstrup walcher certificate of linearity or a calibration certificate from a testing facility accredited by Germany s National Accreditation Body DAkkS is available upon request...

Page 11: ...arting point setting 0 2 1 Digit for 1kPa measurement range 50kPa or 0 5 1 Digit Ambient conditions medium air all non aggressive gases operating temperature 10 C to 40 C storage temperature 10 C to 7...

Page 12: ...ontact with measurement medium beryllium bronze CuBe2 Araldite CY236 HY988 Mu metal nickel alloy Loctite 242e Brass CuZn39Pb3 Carbonyl iron Aluminum AlCuMgPb AlMg3 KEL FPM fluorinated rubber Silicon t...

Page 13: ...KAL 84 Instruction Manual 13 KAL 84 Instruction Manual 9 Dimension drawing 7100 003554_KAL84 doc 14 12 2015 Me...

Reviews: