2.2.9  Time constants 

Pressing this key toggles between three different time constants (response times) for 

the KAL 84 pressure sensor. A red LED indicates the time constant selected. The 

default time constant (20 ms) is used if no LED appears. 

2.2.10  Pressure transducer dial 

Turning the dial generates pressure—left for negative pressure, right for positive 

pressure. The pressure transducer must be brought to its middle position before 

generating negative pressure. This is done by turning the dial to the right several 

times with the pressure ports open. 

Never force the dial!

 

 

Start-up 

Before plugging in the power adapter, make sure the correct supply voltage is 

available. The rechargeable battery (included) will charge up when the instrument is 

run from the power adapter, even if the KAL 84 is not turned on. Because the battery 

may or may not be fully charged upon delivery, we recommend charging the battery 

before start-up. The KAL 84 will only run on battery power if the power adapter is 

unplugged. 

      If the battery has reached the end of its lifecycle and is no longer usable, it 

must be replaced with a 9 V, IEC 6F22 rechargeable battery. 

The power adapter should never be connected if non-rechargeable batteries 

have been installed, as this could destroy the instrument. 

 

Only use the included power adapter! For replacement please contact the 

manufacturer. 

 

 

Pressing the “0/1” switch places the instrument in operating mode. Please remember 

that it takes roughly 30 minutes for the pressure sensor to warm up. During this 

period, the zero point can fluctuate, particularly if the instrument is subject to changes 

in temperature. The automatic zero-point calibration feature should be activated 

numerous times during this period. 

Summary of Contents for KAL 84

Page 1: ...3554 12 2015 Instruction Manual KAL 84 Calibration Device halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup wal...

Page 2: ...6 2 2 User interfaces 6 2 2 1 The 4 digit LCD 7 2 2 2 ON OFF 7 2 2 3 Backlighting 7 2 2 4 Analog output 7 2 2 5 Units of pressure 7 2 2 6 Pressure ports 7 2 2 7 Purge valves 7 2 2 8 Automatic zero po...

Page 3: ...le throughout the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary materials...

Page 4: ...y technical personnel who are appropriately trained and authorized by the operator of the facility may assemble the instrument and set up its electrical connections The instrument may only be operated...

Page 5: ...Symbols The symbols given below are used throughout this manual to indicate instances when improper operation could result in the following hazards WARNING This warns you of a potential hazard that c...

Page 6: ...strument ideal for use in assembling pressure transducers as it allows the transducers to be calibrated on site Use of the KAL 84 for quality assurance purposes e g as a transfer standard for test equ...

Page 7: ...e This feature allows the operator to toggle between 2 units of pressure e g Pa and mmHg A red LED indicates the units used 2 2 6 Pressure ports These ports are for connecting the test object to the K...

Page 8: ...strument is run from the power adapter even if the KAL 84 is not turned on Because the battery may or may not be fully charged upon delivery we recommend charging the battery before start up The KAL 8...

Page 9: ...A manual pump can quickly generate high levels of pressure which can destroy the instrument Caution is particularly advised when using instruments with low measurement ranges 5 Replacing the battery...

Page 10: ...t least once a year A halstrup walcher certificate of linearity or a calibration certificate from a testing facility accredited by Germany s National Accreditation Body DAkkS is available upon request...

Page 11: ...arting point setting 0 2 1 Digit for 1kPa measurement range 50kPa or 0 5 1 Digit Ambient conditions medium air all non aggressive gases operating temperature 10 C to 40 C storage temperature 10 C to 7...

Page 12: ...ontact with measurement medium beryllium bronze CuBe2 Araldite CY236 HY988 Mu metal nickel alloy Loctite 242e Brass CuZn39Pb3 Carbonyl iron Aluminum AlCuMgPb AlMg3 KEL FPM fluorinated rubber Silicon t...

Page 13: ...KAL 84 Instruction Manual 13 KAL 84 Instruction Manual 9 Dimension drawing 7100 003554_KAL84 doc 14 12 2015 Me...

Reviews: