Safety precautions 

1.1 

Appropriate use 

The KAL 84 calibration device is used for testing and calibrating pressure sensors. 

 

Always observe the operating requirements—particularly the permissible supply 

voltage—indicated on the rating plate and in the ”Technical data” section of this 

manual. 

 

Use only an IEC 6F22 rechargeable battery (9 V). Do not use non-rechargeable 

batteries as these can destroy the instrument.

 

 

The instrument may only be handled as indicated in this manual. Modifications to the 

instrument are prohibited. The manufacturer is not liable for damages caused by 

improper use or failure to follow these instructions. Violations of this type render all 

warranty claims null and void.

 

 

 

1.2 

Shipping, assembly, electrical connections and startup 

Do not close the pressure inputs when shipping, as changes in barometric pressure 

could damage instruments with low measuring ranges.

 

 

Only technical personnel who are appropriately trained and authorized by the 

operator of the facility may assemble the instrument and set up its electrical 

connections.

 

 

The instrument may only be operated by appropriately trained individuals who have 

been authorized by the operator of the facility. 

 

Pressurized air or breath is not to be used for performance tests, as this could 

damage instruments with low measurement ranges.

 

 

Measurement errors may occur if the instrument is not kept protected from sunlight. 

 

Specific safety precautions are given in individual sections of this manual. 

 

 

1.3 

Troubleshooting, maintenance, repairs, disposal 

The individual responsible for the electrical connections must be notified immediately 

if the instrument is damaged or if errors occur that cannot be corrected as indicated in 

section 7. 

 

This individual must take the instrument out of service until the error has been 

corrected and ensure that it cannot be used unintentionally. 

 

Always unplug the power cord before opening the instrument! 

 

This instrument requires no maintenance. 

 

Only the manufacturer may perform repairs that require the housing to be opened. 

Summary of Contents for KAL 84

Page 1: ...3554 12 2015 Instruction Manual KAL 84 Calibration Device halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup wal...

Page 2: ...6 2 2 User interfaces 6 2 2 1 The 4 digit LCD 7 2 2 2 ON OFF 7 2 2 3 Backlighting 7 2 2 4 Analog output 7 2 2 5 Units of pressure 7 2 2 6 Pressure ports 7 2 2 7 Purge valves 7 2 2 8 Automatic zero po...

Page 3: ...le throughout the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary materials...

Page 4: ...y technical personnel who are appropriately trained and authorized by the operator of the facility may assemble the instrument and set up its electrical connections The instrument may only be operated...

Page 5: ...Symbols The symbols given below are used throughout this manual to indicate instances when improper operation could result in the following hazards WARNING This warns you of a potential hazard that c...

Page 6: ...strument ideal for use in assembling pressure transducers as it allows the transducers to be calibrated on site Use of the KAL 84 for quality assurance purposes e g as a transfer standard for test equ...

Page 7: ...e This feature allows the operator to toggle between 2 units of pressure e g Pa and mmHg A red LED indicates the units used 2 2 6 Pressure ports These ports are for connecting the test object to the K...

Page 8: ...strument is run from the power adapter even if the KAL 84 is not turned on Because the battery may or may not be fully charged upon delivery we recommend charging the battery before start up The KAL 8...

Page 9: ...A manual pump can quickly generate high levels of pressure which can destroy the instrument Caution is particularly advised when using instruments with low measurement ranges 5 Replacing the battery...

Page 10: ...t least once a year A halstrup walcher certificate of linearity or a calibration certificate from a testing facility accredited by Germany s National Accreditation Body DAkkS is available upon request...

Page 11: ...arting point setting 0 2 1 Digit for 1kPa measurement range 50kPa or 0 5 1 Digit Ambient conditions medium air all non aggressive gases operating temperature 10 C to 40 C storage temperature 10 C to 7...

Page 12: ...ontact with measurement medium beryllium bronze CuBe2 Araldite CY236 HY988 Mu metal nickel alloy Loctite 242e Brass CuZn39Pb3 Carbonyl iron Aluminum AlCuMgPb AlMg3 KEL FPM fluorinated rubber Silicon t...

Page 13: ...KAL 84 Instruction Manual 13 KAL 84 Instruction Manual 9 Dimension drawing 7100 003554_KAL84 doc 14 12 2015 Me...

Reviews: