KAL 84 Instruction Manual
KAL 84 Instruction Manual
10
6 Calibration
Please keep in mind that part I of DIN 16 005 requires reference instruments to be at
least four times as precise as the device being calibrated.
Equipment used to test halstrup-walcher pressure calibration devices are monitored
continuously and our instruments are themselves calibrated using reference
pressures based on national standards.
In order to maintain this high level of quality over a long period of time, we
recommend sending your pressure calibration device to the manufacturer for
recalibration at least once a year. A halstrup-walcher certificate of linearity or a
calibration certificate from a testing facility accredited by Germany's National
Accreditation Body (DAkkS) is available upon request.
Please note when the next calibration is due (indicated on the rear of the
instrument).
7
Troubleshooting
Problem
Cause
Corrective Action
KAL 84 will not switch on
power adapter is not plugged in
(to electrical outlet)
power adapter is not plugged in
(to the KAL 84)
incorrect supply voltage
dead battery
plug the adapter into the
outlet
connect the adapter to the
KAL 84
provide correct supply voltage
recharge or replace battery
Set pressure does not
remain constant
purge valve is open
leak in tubing
low battery
close purge valve
inspect tubing
recharge or replace battery
Table 2