Troubleshooting

The following table of potential indicators provides a quick reference of the descriptions, probable causes 
and most common solutions regarding the operation of the HaloFogger®. If the information in this table 
does not describe or solve the problem you are having, contact Customer Service at 1 302-454-8102, 
Monday through Friday, 9:00AM to 5:00PM EST. 

Symptom

Possible Cause

Solution

No mist is being 
produced

The nozzle is clogged 
misaligned.

Remove the nozzle and clean according to 
the instructions in this manual.

The disinfectant fluid
tank is empty.

Check the fluid level indicator lights. If the 
orange light is illuminated, follow the fill 
instructions in this manual to refill the tank.

Start button not pressed.

Press and hold the Start button for 2 
seconds to begin fogging.

Green indicator light is 
blinking

Reservoir Door is not  
closed correctly.

Close Reservoir Door making sure it “clicks”  
when closed.

Magnet in Reservoir Door 
is not in place.  

Call Customer Service.

No power 

The power cord is not 
plugged in or the power 
source is not on.

Check to make sure that the power cord is 
plugged into both the fogger and the wall 
outlet. Then, check to see that the outlet is 
working. The green status indicator light 
should be illuminated when properly set up. 
 

The fuse is blown.

Check to see if the fuse is blown. Fuse (and 
a spare) is located behind a discreet panel 
just below where the power cord connects 
to the HaloFogger. 

Knocking noise

The compressor is not 
running properly.

Call Customer Service.

Disinfecting procedure 
is interrupted

Power loss during 
treatment process 
or other unforeseen 
circumstances

.

Confirm power has been restored. Re-
enter room after hydrogen peroxide 
concentration level is determined to be at 
or below 1 ppm. Restart treatment process 
from beginning. 

HaloFogger® Operation Manual_1911

The unit not plugged in.

Check to make sure that the power cord 
is plugged into both the HaloFogger and 
a wall outlet. Then, check to see that the 
wall outlet is working. The green status 
indicator light should be illuminated when 
properly set up.

Summary of Contents for HaloFogger STD 1911

Page 1: ...READ AND SAVE THESE INSTRUCTIONS Operation Manual 110V 220V STD 1911...

Page 2: ...treatment process Reservoir Door Funnel Port Status Indicator Light Walk Around Tour of the HaloFogger Status Indicator Light Displays the operational status of the Halo device Solid Green Indicates t...

Page 3: ...s when pouring disinfectant fluid into the device Avoid splashing or overfilling which could harm the user and or the device 12 Halosil International is not responsible for any injury or damage caused...

Page 4: ...pen a gallon 3 8 liters bottle of HaloMist Disinfectant and SLOWLY pour the contents into the funnel area AVOID SPLASHING 5 The fluid level indicator lights will illuminate as you pour to show how muc...

Page 5: ...ensor assume a waiting time of AT LEAST 110 MINUTES AFTER FOGGING 110 minutes is the minimum wait time Length of wait time increases with room size See time chart below Minimum Wait Times Fog Time Rec...

Page 6: ...om is not blocked c Position the dehumidifier if one is being used outside the smaller interior room centered in the doorway facing towards the fogger d Use Chemical Indicator Strips to ensure that th...

Page 7: ...d can be connected to the HaloFogger data logger via Bluetooth The HaloLog App does not need to be connected to a fogger before a treatment run is started The on board data logger captures all of the...

Page 8: ...will be displayed 4 To initiate a connection tap on any HaloFogger serial number displayed on the Halo List Screen When the connection is made the Run Events Screen will appear The top row of the scre...

Page 9: ...er and total Run Time for that HaloFogger highlighted in GREEN The HaloLog data stored on the selected HaloFogger will download automatically after the Bluetooth handshake has been established which i...

Page 10: ...vents will identified BILLABLE as a default Events that are not treatments a test run for example can be identified as NON BILLABLE These events will not appear in the downloadable CSV file but will b...

Page 11: ...address A keyboard will appear to enter address ACTION Tap to email data The iDevice s email screen will open to confirm message ACTION Tap SEND in upper right corner of screen to send message Back u...

Page 12: ...me to a HaloFogger 2 In the event that HaloLogs of the same HaloFogger have been downloaded to multiple devices If one user has made changes to the HaloLog for example added room building or operator...

Page 13: ...remove Step 3 Separate Parts Remove Main Bolt from Nozzle Insert Soak both parts in soap and water Rinse parts thoroughly with water Step 4 Reassemble Nozzle Carefully insert the Main Bolt back into...

Page 14: ...Reservoir Door is not in place Call Customer Service No power The power cord is not plugged in or the power source is not on Check to make sure that the power cord is plugged into both the fogger and...

Page 15: ...Nozzle Assembly 0750 000 includes retainer cap nozzle insert main bolt and O ring Re Order Number Part Part Number 110V HaloFogger US and Canada 0753 110 220V HaloFogger 0753 220 HaloMist All Purpose...

Page 16: ...of room being treated CAUTION PRECAUCION DO NOT ENTER ENTERINO Room Fogging in Progress Treatment Date Start Time Estimated Room Re entry Time Refer to HaloMist Use Instructions to determine room re...

Page 17: ...ion phone and email address purchase date from which HaloFogger was purchased and the unit s serial number Proof of purchase may be required If Halosil International determines that the fogger qualifi...

Page 18: ...uce infection rates at your facility by visiting www halosil com There you can find out more about the Halo Disinfection System download literature and case studies and find a distributor near you 91...

Reviews: