23
Control
4205
R
The maximum flow measured of the current, or most recent
measurement, a signed 2-byte integer. Flow is equal to the
value after formatting with the factory-set number of decimal
places. This value will not be accurate if flow range, units, gas, or
calibration has been changed.
Added in V3.0.0.
4206
R
The number of samples averaged in the current, or most recent
measurement, an unsigned integer. This number will not be
larger than the last value written to register 4201.
Added in V3.0.0.
4207
R
The average temperature of the current, or most recent
measurement, a signed integer. Temperature (°C) = value / 100.
Added in V3.0.0.
4208
R
The average flow of the current, or most recent measurement, a
signed 2-byte integer. Flow is equal to the value after formatting
with the factory-set number of decimal places. This value will
not be accurate if flow range, units, gas, or calibration has been
changed.
Added in V3.0.0.
4209
R
The instantaneous valve drive. Valve drive (%) = value / 100.
Added in V3.0.0.
4210
R
The minimum flow measured in the previously completed
measurement, a signed 2-byte integer. Flow is equal to the
value after formatting with the factory-set number of decimal
places. This value will not be accurate if flow range, units, gas, or
calibration has been changed.
Added in V3.0.0.
4211
R
The maximum flow measured in the previously completed
measurement, a signed 2-byte integer. Flow is equal to the
value after formatting with the factory-set number of decimal
places. This value will not be accurate if flow range, units, gas, or
calibration has been changed.
Added in V3.0.0.
4212
R
The number of samples averaged in the previously completed
measurement, an unsigned integer.
Added in V3.0.0.
4213
R
The average temperature of the previously completed
measurement, a signed integer. Temperature (°C) = value / 100.
Added in V3.0.0.
4214
R
The average flow of the previously completed measurement, a
signed 4-byte integer. Flow is equal to the value after formatting
with the factory-set number of decimal places. This value will
not be accurate if flow range, units, gas, or calibration has been
changed.
Added in V3.0.0.