9
Control
Note that the following three controls differ from the FlowVision manual for BASIS 2:
1.
Valve drive. The “Valve” button has the following options:
a. Open/exhaust
b. Close
c. Direct drive (sets the "Setpoint" entry field to use a percentage instead of a unit of
measure)
d. Resume (clears the direct drive and returns to closed-loop control).
2.
The totalizer limit mode, shown in
Instrument
Settings > Totalizer
, is unique to
BASIS 2
3.
Setpoint ramping. The button near the “Setpoint” box will read Ramp: (disabled) or (%/
msec). Clicking the button will open
Instrument Settings > Control
, where a new
ramp rate can be set or cleared
.