25
Control
49
R
The units of measure for flow:
0: SCCM
1: NCCM
2: SLPM
3: NLPM
4: SmL/s
5: NmL/s
6: SmL/m
7: NmL/m
8: SL/h
9: NL/h
10: SCCS
11: NCCS
12: Sm³/h
13: Nm³/h
14: Sm³/d
15: Nm³/d
16: SCIM
17: SCFM
18: SCFH
19: SCFD
Added more units in V3.0.0.
51
R/W
The number of samples to use when performing a flow tare, an
integer 1–65535. Each sample is taken at approximately a 2.5
msec interval.
Added in V2.4.0.
52
R/W
Adjust the reported flow values to this reference temperature,
an integer 0–3000. Reference temperature (°C) = value / 100.
Added in V2.4.0.
53
W
Write 0xAA55 (43605) to reset the total volume to 0.
Added in V2.5.5.
54
R/W
The totalizer behavior when the total volume reaches the
maximum limit the instrument is capable of, represented as an
integer:
0: Keep total volume at max value and do not indicate OVR error.
1: Reset total volume to zero and do not indicate an OVR error.
2: Keep total volume at max value and indicate an OVR error.
3: Reset total volume to zero and indicate an OVR error.
Added in V3.0.0.
55
R/W
The averaging to use for the reported flow. An integer 0-2500.
The time constant is approximately value milliseconds.
Added in V3.0.0.