Hallowell EMC 2002PRO Operating Manual Download Page 16

2/26/2018 

DOCA3667B Pro Series Operating Manual.doc 

16 of 22 

OPERATING INSTRUCTIONS 

 

 

 

 

 

 

 

 
Output of the 

Model 2002

PRO 

/2002IE

PRO

 

is adjusted by only two controls; a linearly 

calibrated RATE control (breaths per minute), and a metering VOLUME control. 
 
Follow the setup and verification procedures and be certain the pop-off valve on the anesthesia 
machine is completely closed. 

DETAILED OPERATING INSTRUCTIONS: 

 
1.

 

When the patient is ready, reconnect the ventilator to breathing system. 

 
2.

 

ALWAYS set the VOLUME controls to its minimum setting before turning on the ventilator.  
The VOLUME controls must be fully clockwise to be set to the minimum setting. 

 
3.

 

Set the maximum working pressure limit (MWPL).   

 
4.

 

Turn on the ventilator. Set the RATE before any adjustment is made to the VOLUME.  
Adjust the RATE to the desired breaths per minute setting. 

 

NOTE:  As you proceed, continually observe the anesthesia system's breathing system pressure 

(BSP) gauge to ensure that excessive pressures are not being attained. 

 
5.

 

The volume controls, as noted above, are initially set after the appropriate RATE has been 
selected.  Turn the VOLUME controls counterclockwise to increase the tidal volume 
delivered, or clockwise, to decrease the tidal volume delivered.  Read the approximate tidal 
volume by noting the displacement of the bellows in ml as indicated on the bellows housing 
scale.  At a given RATE setting, the tidal volume can be increased or decreased in this 
manner. 

 
6.

 

Slight changes in he RATE and VOLUME controls can be made as the procedure continues 
but never make any gross adjustments to these controls with the patient connected. 

 

 

Warning: 

Under no circumstances should the flush button on the anesthesia machine be used during the 
Inspiratory phase of the breathing cycle.  There is the extreme danger of rupturing a lung.  The 
flush button introduces 50 - 100 lpm, perhaps more, of oxygen flow into the breathing system.  
During inspiration the discharge valve in the control unit is closed so that flush flow is added to 
the inspiratory flow generated by the ventilator and has no where else to go except to the patient's 
lungs. It is recommended that the flush feature on the anesthesia machine NEVER be used with 
patient connected.  The oxygen flow valve can be opened further than normal providing a more 
controllable high flow of oxygen. 

 

Summary of Contents for 2002PRO

Page 1: ...Hallowell EMC Model 2002PRO Model 2002IEPRO Veterinary Anesthesia Ventilators OPERATING SET UP MANUAL...

Page 2: ...CAUTIONS 3 WARRANTY 4 5 INTRODUCTION 6 RECEIVING PROCEDURES 7 SET UP PROCEDURE SET UP DIAGRAM 8 9 10 CONTROLS 11 ALARMS 12 VERIFICATION OF PROPER FUNCTION 12 13 GETTING STARTED 14 15 OPERATING INSTRU...

Page 3: ...ons unauthorized repair or modification of the equipment or accessories or from the use of components or accessories that have either been damaged or not authorized for use with this equipment by Hall...

Page 4: ...ited warranty WHAT THIS WARRANTY DOES NOT COVER 1 Defects or damages resulting from use of the product in other than its normal and customary manner 2 Defects or damages from abnormal use abnormal con...

Page 5: ...low the exclusive of limitation of incidental or consequential damages or limitations on how long an implied warranty lasts these limitations or exclusions may not apply to you This warranty gives you...

Page 6: ...impact resistant material that is unaffected by water or the presence of all state of the art anesthetics In addition the high temperature properties of ULTEM make it fully compatible with standard s...

Page 7: ...ng Model 2002PRO 2002IEPRO Controller Bellows Base Assembly PN 000A0510 Tube Driving Gas 7 black rubber PN 000A0495 Airway Pressure Sampling Tee PN 000A2420B 36 x 22 mm Breathing System Tube PN 201A16...

Page 8: ...unting ring Carefully hold the outer edge of the bellows disk top of bellows lift and lower it quickly several times to puff out and remove any folds in the convolutions Place the bellows housing over...

Page 9: ...r is very economical therefore the use of oxygen as a drive gas is recommended Oxygen use reduces the risk of unit malfunction due to contamination of the pneumatics Compressed air may be used but it...

Page 10: ...2 26 2018 DOCA3667B Pro Series Operating Manual doc 10 of 22...

Page 11: ...ntrol is described in detail below I O Power Switch a toggle switch used to turn the ventilator on The green LED indicates the power is on when illuminated INSPiratory HOLD Control A momentary pushbut...

Page 12: ...sor for this alarm is located downstream of the internal 40 micron filter and may also indicate a clogged filter condition During an Alarm condition the yellow light on the front panel is illuminated...

Page 13: ...the breathing system is not leaking The pressure should remain constant Release the INSPiratory HOLD button Observe that the ventilator is continuing to cycle and that over time ten cycles and the bel...

Page 14: ...the maximum working pressure limit MWPL control to about 20 cm H2O The airway pressure will not exceed this setting regardless of what you do with the other controls Connect the ventilator to the bre...

Page 15: ...ge being implemented you have changed the MV as desired now If needed trim the delivered TV size with the RATE control to obtain a new TV that results in a more appropriate PIP and no alarms Now do it...

Page 16: ...ppropriate RATE has been selected Turn the VOLUME controls counterclockwise to increase the tidal volume delivered or clockwise to decrease the tidal volume delivered Read the approximate tidal volume...

Page 17: ...a tight o ring fit Tilt the top of the bellows housing toward you and lift it off DO NOT attempt to steam sterilize the bellows housing Since it does not come in contact with the breathing gas it nee...

Page 18: ...the base and loosen the four black thumbscrews located at the corners First clean the base using a clean soft slightly dampened cloth Then wrap the whole base and steam sterilize it using the same sta...

Page 19: ...produce a closed loop 4 While observing the breathing system pressure slowly increase the fresh gas flow to the system until the pressure builds to about 50 cm H2O 5 Turn the flow off 6 The pressure s...

Page 20: ...llapsed bellows Unkink the drive gas tube Unkink breathing system hose The ventilator operation sounds normal but the TV delivered is incorrect and or inconsistent Missing or damaged pop off valve o r...

Page 21: ...name address phone number and PO number of the responsible party 5 If the item is to be returned such as with a repair the name of the party to whom we should ship the item the shipping address and r...

Page 22: ...mum Working Pressure Limit Linearly Adjustable 10 60 cm H2O Power On Front Panel mounted green LED Alarm Visual Front Panel mounted yellow LED Alarm Audio Internal audio transducer Alarm Audio Int Low...

Reviews: