Hallowell EMC 2002PRO Operating Manual Download Page 12

2/26/2018 

DOCA3667B Pro Series Operating Manual.doc 

12 of 22 

ALARMS 

 

 

 

 

 

 

 

 

 

 

 

 

Maximum Working Pressure Limit (MWPL):

  The MWPL feature allows the operator to set 

an upper limit above which the airway pressure will not exceed.  The ventilator will terminate the 
inspiratory phase of the breathing cycle and begin an expiratory phase when the pressure 
transducer senses a pressure above the MWPL setting.  The MWPL is settable over a range from 
10 to 60 cm H

2

O.  When the airway pressure reaches the set limit, the yellow light on the front 

panel blinks and a short tone is heard.  If the excessive pressure is not immediately relieved, 
cycling is paused and the alarm sounds continuously.  Note that the INSPiratory HOLD feature is 
designed to not function when the MWPL setting is exceeded.  Therefore, a holding inspiration 
will be released when the MWPL setting is exceeded. 
 

Low Breathing System Pressure Alarm (LO BSP):

  The LO BSP alarm is activated at the end 

of inspiration if there is not at least 5 cm H

2

O pressure sensed by the pressure transducer.  This 

alarm is sometimes commonly referred to as a "disconnect" alarm; however, it should be 
understood that a patient disconnect is not always nor the only cause of low breathing system 
pressure.  During the alarm condition a yellow light is illuminated on the front panel and the 
sound of a raspy siren is heard.  The alarm is automatically reset at the end of the next

 

inspiratory 

phase in which there is a minimum of 5 cm H

2

O pressure sensed by the pressure transducer. 

 
Low 50 psi Supply Gas Alarm:

  The Low Supply Gas alarm is activated when the supply gas 

pressure drops below 35 psi (2.4 bar).  The sensor for this alarm is located downstream of the 
internal 40-micron filter and may also indicate a clogged filter condition.  During an Alarm 
condition, the yellow light on the front panel is illuminated and a steady, continuous tone is 
heard.  This alarm automatically resets when the pressure increases above 40 psi (2.7 bar). 
 

VERIFICATION OF PROPER FUNCTION 

 

 

 

 

 

 
 

A Note on Test Lungs 

The most readily available test lung will probably be the breathing bag you removed to connect 
the ventilator.  A breathing bag is a very poor model of a lung.  It can be used if one understands 
how poor a model of the lung it is and how to avoid using it in such a way that it creates problems 
that will not occur with a real patient. 
 
A far better test lung is a ridged walled container such as a gas can, water fountain bottle, or beer 
keg.  The compliance of these containers is equal to their volume in liters and will be linear as is 
that of a real lung over normal operating ranges.  Most importantly they will maintain a 
functional residual capacity (FRC) that is hard to maintain in a breathing bag. 
 
The breathing bag has an unpredictable FRC from breath to breath if bumped or squeezed beyond 
the point of relaxation, more gas than would normally be popped off at the end of exhalation 
escapes from the breathing system.  The bellows then abnormally fails to remain at the top of the 
bellows housing at the end of exhalation.  To use the bag successfully, connect it to the patient 
wye, hang it vertically, and do not disturb it. 

 
 
Verification of proper ventilator operation requires that you first complete the system setup, as 
described on page 7 and connect a test lung to the patient wye piece.  During verification, you 
will be observing the operation of the entire system configuration, checking for leaks, and 
monitoring the ventilator for consistent cycling. 
 

Summary of Contents for 2002PRO

Page 1: ...Hallowell EMC Model 2002PRO Model 2002IEPRO Veterinary Anesthesia Ventilators OPERATING SET UP MANUAL...

Page 2: ...CAUTIONS 3 WARRANTY 4 5 INTRODUCTION 6 RECEIVING PROCEDURES 7 SET UP PROCEDURE SET UP DIAGRAM 8 9 10 CONTROLS 11 ALARMS 12 VERIFICATION OF PROPER FUNCTION 12 13 GETTING STARTED 14 15 OPERATING INSTRU...

Page 3: ...ons unauthorized repair or modification of the equipment or accessories or from the use of components or accessories that have either been damaged or not authorized for use with this equipment by Hall...

Page 4: ...ited warranty WHAT THIS WARRANTY DOES NOT COVER 1 Defects or damages resulting from use of the product in other than its normal and customary manner 2 Defects or damages from abnormal use abnormal con...

Page 5: ...low the exclusive of limitation of incidental or consequential damages or limitations on how long an implied warranty lasts these limitations or exclusions may not apply to you This warranty gives you...

Page 6: ...impact resistant material that is unaffected by water or the presence of all state of the art anesthetics In addition the high temperature properties of ULTEM make it fully compatible with standard s...

Page 7: ...ng Model 2002PRO 2002IEPRO Controller Bellows Base Assembly PN 000A0510 Tube Driving Gas 7 black rubber PN 000A0495 Airway Pressure Sampling Tee PN 000A2420B 36 x 22 mm Breathing System Tube PN 201A16...

Page 8: ...unting ring Carefully hold the outer edge of the bellows disk top of bellows lift and lower it quickly several times to puff out and remove any folds in the convolutions Place the bellows housing over...

Page 9: ...r is very economical therefore the use of oxygen as a drive gas is recommended Oxygen use reduces the risk of unit malfunction due to contamination of the pneumatics Compressed air may be used but it...

Page 10: ...2 26 2018 DOCA3667B Pro Series Operating Manual doc 10 of 22...

Page 11: ...ntrol is described in detail below I O Power Switch a toggle switch used to turn the ventilator on The green LED indicates the power is on when illuminated INSPiratory HOLD Control A momentary pushbut...

Page 12: ...sor for this alarm is located downstream of the internal 40 micron filter and may also indicate a clogged filter condition During an Alarm condition the yellow light on the front panel is illuminated...

Page 13: ...the breathing system is not leaking The pressure should remain constant Release the INSPiratory HOLD button Observe that the ventilator is continuing to cycle and that over time ten cycles and the bel...

Page 14: ...the maximum working pressure limit MWPL control to about 20 cm H2O The airway pressure will not exceed this setting regardless of what you do with the other controls Connect the ventilator to the bre...

Page 15: ...ge being implemented you have changed the MV as desired now If needed trim the delivered TV size with the RATE control to obtain a new TV that results in a more appropriate PIP and no alarms Now do it...

Page 16: ...ppropriate RATE has been selected Turn the VOLUME controls counterclockwise to increase the tidal volume delivered or clockwise to decrease the tidal volume delivered Read the approximate tidal volume...

Page 17: ...a tight o ring fit Tilt the top of the bellows housing toward you and lift it off DO NOT attempt to steam sterilize the bellows housing Since it does not come in contact with the breathing gas it nee...

Page 18: ...the base and loosen the four black thumbscrews located at the corners First clean the base using a clean soft slightly dampened cloth Then wrap the whole base and steam sterilize it using the same sta...

Page 19: ...produce a closed loop 4 While observing the breathing system pressure slowly increase the fresh gas flow to the system until the pressure builds to about 50 cm H2O 5 Turn the flow off 6 The pressure s...

Page 20: ...llapsed bellows Unkink the drive gas tube Unkink breathing system hose The ventilator operation sounds normal but the TV delivered is incorrect and or inconsistent Missing or damaged pop off valve o r...

Page 21: ...name address phone number and PO number of the responsible party 5 If the item is to be returned such as with a repair the name of the party to whom we should ship the item the shipping address and r...

Page 22: ...mum Working Pressure Limit Linearly Adjustable 10 60 cm H2O Power On Front Panel mounted green LED Alarm Visual Front Panel mounted yellow LED Alarm Audio Internal audio transducer Alarm Audio Int Low...

Reviews: