Hallowell EMC 2002PRO Operating Manual Download Page 15

2/26/2018 

DOCA3667B Pro Series Operating Manual.doc 

15 of 22 

GETTING STARTED (Continued) 

Our ventilator, for those of you familiar with the term, is considered a MV divider. For those 
unfamiliar with the term, the ventilator delivers a consistent MV to the patient and that MV is 
divided into different size TVs by the RATE control.  You can change the RATE control all you 
want without changing the total ventilation delivered to the patient.  Let me repeat that.  You can 
change the RATE control all you want and it will not effect the total ventilation delivered to the 
patient.**  In order to change the MV delivered you need only, in fact you must, change the 
VOLUME controls.  This point will be quite important when you go to wean the patient from the 
ventilator. 

Remember, when you change the VOLUME controls it is the inspiratory flow that you are 
changing directly.  The rate and, therefore, the time that flow is delivered has not changed, thus 
the delivered TV will be either larger or smaller than before.  Stop and think about it, you are now 
delivering a different volume of gas to the same compliance of the patient; it follows that the PIP 
will be different.  This different PIP may be fine or it may be unnecessarily high or low.  In the 
high extreme the MWPL alarm will sound, a short steady tone, and the PIP will be limited to the 
set value or in the low extreme, the LO BSP alarm will sound, a warbling tone. 

Back to the change being implemented, you have changed the MV as desired now, If needed, trim 
the delivered TV size with the RATE control to obtain a new TV that results in a more 
appropriate PIP, and no alarms. 

Now do it and get comfortable with it.  

Don't just put this document away - setup the ventilator 

and a test lung.  Read this again trying what is discussed as you read. 

Note that no discussion has been made of I-time and E-time and the need to keep them in a proper 
relation to each other.  This relation is automatically held constant by the ventilator.  The I:E ratio 
is a consistent 1:2, no need to think about it - there will be enough time for exhalation.  For those 
of you that want to think about it, we offer the model 2KIE with an adjustable ratio from 1:1.5 to 
1:4.  Even when the I:E ratio is adjustable that ratio is still held constant over the full range of rate 
settings. 

**This statement is somewhat of a simplification as you deviate greatly from the current RATE 
setting.  There is a difference in MV delivered by the ventilator and the alveolar ventilation 
received by the patient.  This difference is related to the dead space and BS compliance.  With 
each TV delivered, a portion ventilates the dead space and BS, the more TVs per minute the 
greater the portion of the delivered MV that is not seen by the alveoli and, thus, is of no use to the 
patient. The significance of this difference is small unless the BS being used is severely 
mismatched with the patient or the deviation from the current setting is great.  Similarly the 
amount of variation is minimal with the small changes needed to trim the TV after adjusting the 
MV. 

Summary of Contents for 2002PRO

Page 1: ...Hallowell EMC Model 2002PRO Model 2002IEPRO Veterinary Anesthesia Ventilators OPERATING SET UP MANUAL...

Page 2: ...CAUTIONS 3 WARRANTY 4 5 INTRODUCTION 6 RECEIVING PROCEDURES 7 SET UP PROCEDURE SET UP DIAGRAM 8 9 10 CONTROLS 11 ALARMS 12 VERIFICATION OF PROPER FUNCTION 12 13 GETTING STARTED 14 15 OPERATING INSTRU...

Page 3: ...ons unauthorized repair or modification of the equipment or accessories or from the use of components or accessories that have either been damaged or not authorized for use with this equipment by Hall...

Page 4: ...ited warranty WHAT THIS WARRANTY DOES NOT COVER 1 Defects or damages resulting from use of the product in other than its normal and customary manner 2 Defects or damages from abnormal use abnormal con...

Page 5: ...low the exclusive of limitation of incidental or consequential damages or limitations on how long an implied warranty lasts these limitations or exclusions may not apply to you This warranty gives you...

Page 6: ...impact resistant material that is unaffected by water or the presence of all state of the art anesthetics In addition the high temperature properties of ULTEM make it fully compatible with standard s...

Page 7: ...ng Model 2002PRO 2002IEPRO Controller Bellows Base Assembly PN 000A0510 Tube Driving Gas 7 black rubber PN 000A0495 Airway Pressure Sampling Tee PN 000A2420B 36 x 22 mm Breathing System Tube PN 201A16...

Page 8: ...unting ring Carefully hold the outer edge of the bellows disk top of bellows lift and lower it quickly several times to puff out and remove any folds in the convolutions Place the bellows housing over...

Page 9: ...r is very economical therefore the use of oxygen as a drive gas is recommended Oxygen use reduces the risk of unit malfunction due to contamination of the pneumatics Compressed air may be used but it...

Page 10: ...2 26 2018 DOCA3667B Pro Series Operating Manual doc 10 of 22...

Page 11: ...ntrol is described in detail below I O Power Switch a toggle switch used to turn the ventilator on The green LED indicates the power is on when illuminated INSPiratory HOLD Control A momentary pushbut...

Page 12: ...sor for this alarm is located downstream of the internal 40 micron filter and may also indicate a clogged filter condition During an Alarm condition the yellow light on the front panel is illuminated...

Page 13: ...the breathing system is not leaking The pressure should remain constant Release the INSPiratory HOLD button Observe that the ventilator is continuing to cycle and that over time ten cycles and the bel...

Page 14: ...the maximum working pressure limit MWPL control to about 20 cm H2O The airway pressure will not exceed this setting regardless of what you do with the other controls Connect the ventilator to the bre...

Page 15: ...ge being implemented you have changed the MV as desired now If needed trim the delivered TV size with the RATE control to obtain a new TV that results in a more appropriate PIP and no alarms Now do it...

Page 16: ...ppropriate RATE has been selected Turn the VOLUME controls counterclockwise to increase the tidal volume delivered or clockwise to decrease the tidal volume delivered Read the approximate tidal volume...

Page 17: ...a tight o ring fit Tilt the top of the bellows housing toward you and lift it off DO NOT attempt to steam sterilize the bellows housing Since it does not come in contact with the breathing gas it nee...

Page 18: ...the base and loosen the four black thumbscrews located at the corners First clean the base using a clean soft slightly dampened cloth Then wrap the whole base and steam sterilize it using the same sta...

Page 19: ...produce a closed loop 4 While observing the breathing system pressure slowly increase the fresh gas flow to the system until the pressure builds to about 50 cm H2O 5 Turn the flow off 6 The pressure s...

Page 20: ...llapsed bellows Unkink the drive gas tube Unkink breathing system hose The ventilator operation sounds normal but the TV delivered is incorrect and or inconsistent Missing or damaged pop off valve o r...

Page 21: ...name address phone number and PO number of the responsible party 5 If the item is to be returned such as with a repair the name of the party to whom we should ship the item the shipping address and r...

Page 22: ...mum Working Pressure Limit Linearly Adjustable 10 60 cm H2O Power On Front Panel mounted green LED Alarm Visual Front Panel mounted yellow LED Alarm Audio Internal audio transducer Alarm Audio Int Low...

Reviews: