34
Name
Blood Bank Refrigerator
Model
HXC-358
HXC-358B
Exterior dimensions
(W x D x H) (mm)
720x690x1730
Interior dimensions
(W x D x H) (mm)
620x490x1160
Eff ective volume
358L
Door
Glass door with electric heat
Insulation
CFC-Free foamed-in-place urethane
Compressor
High Quality Hermetically sealed compressor
Shelves / Drawer
5 shelves / 20 baskets
5 Drawers
Load
Max:25kg per shelf
Refrigerating Method
Forced convection cooling
Exterior / Interior
Scratch resistant painted steel / Stainless steel
Condenser/Evaporater
Fin tube type
Temperature Control
Microprocessor controlled
Refrigerant
R134a 420g
Interior light
LED 3W
Net weight
158kg
165kg
Temperature range
4±1
℃
Voltage
220V-240V~/50Hz/60Hz
Rated power / current
460W/3.0A
Noise Level
43dB(A)
Foaming Agent
HFO-1233zde
Climate Type
ST
Anti-shock Safety
Classifi cation
I
Power connection type
Y
USB function / temperature
recorder
Optional / Standard
Alarm system
High-low temperature alarm, sensor fault alarm, power failure alarm,
low battery alarm and door opening alarm
Battery duration for alarm
system
48 hours (when battery is fully charged)
Rechargeable battery
DC12V Rechargeable lead-acid battery
Note: Climate Type ST means the temperature is 18 - 38
℃
,the relative humidity is up to 85%.Due
to continuous improvements, technical information on your refrigerator may be diff erent from herein
published.
7
•Ambient temperature: 18
℃
to 38
℃
with exception of HXC-106/1308/1308B at 16
℃
to 32
℃
. Range
of 18
℃
to 25
℃
is ideal.
•Ambient humidity: below 85% RH .
•Avoid dusty and unstable areas for installation.
•Avoid mechanical swing or vibration.
•The refrigerator shall operate at an altitude lower than 2000m.
•Input voltage: Within rated voltage ±10%.
The installation site shall meet following requirements for normal operation and best performance of
refrigerator:
•Do not install the refrigerator in a narrow and confi ned space as heat may be trapped to increase
ambient temperature. Additionally, a unit should not be installed in a space where the entry door way
is low to prevent an easy maneuver of the unit for maintenance and operation.
•The ground at the installation site shall be solid and fl at.
•The fl oor for the installation site should be solid and fl at. It should also be well ventilated and free
from direct sunshine.
•The power outlet for the refrigerator must be a dedicated power outlet. The power plug must be
plugged in to the outlet securely.
•Do not twist or jam the power cord.
•If the power cord needs to be extended, the cross section area of the conductor in the extended line
should be no less than 3 squared millimeters and no longer than 3 meter.
•Inspect the operating voltage range before operation. If the voltage supply is unstable, install a
voltage stabilizer of 4000W or greater to ensure the supply voltage is within 10% of the rated voltage
supply.
•The refrigerator must be grounded reliably.
•Check the integrity of the socket ground before power up. If it is not grounded appropriately, repair
the wiring before installing the unit.
•If the socket is not provided with grounding wire, a new one shall be installed by engineering
technician.
•Refrigerator units are not permitted to operate outdoors. Rough condition and
rainwater can cause electrical issues and hazardous conditions.
•It is prohibited to install the refrigerator outdoors. Electric leakage or shock may be
caused if the refrigerator gets wet by rainwater.
•Do not ground the refrigerator through gas lines, water lines, telephone lines, or
lighting rods as these devices may cause electric shock and danger.
•The power plug and outlet should be located in a place where it can be accessed
easily and immediately in case of an emergency. Air vents must be free of obstruction.
Product Installation
Installation environment
Caution
Installation site
Warning
Summary of Contents for HXC-106