Haier HXC-106 Operation Manual Download Page 2

Content

Safety Precautions ............................................................ 1

Application guidelines ........................................................ 5

Product Feature ................................................................. 6

Product Installation ............................................................ 7

Component Names●Control Panel  .................................. 12

Application method ........................................................... 16

Display and alarm ............................................................. 22

Cleaning and Maintenance ............................................... 26

FAQ .................................................................................. 28

Refrigeration Schematic • Circuit Diagram ....................... 29

Specifi cation • Packing List .............................................. 33

41

Summary of Contents for HXC-106

Page 1: ......

Page 2: ...guidelines 5 Product Feature 6 Product Installation 7 Component Names Control Panel 12 Application method 16 Display and alarm 22 Cleaning and Maintenance 26 FAQ 28 Refrigeration Schematic Circuit Di...

Page 3: ...en replacing the light from the power source to avoid electric hazards Warning 1 Unauthorized opening of the top cover of the control cabinet is prohibited to avoid damages to the parts inside or inju...

Page 4: ...it is equipped with a 16A or 13A power plug Do not remove the ground pin of the power plug under any circumstances Make sure the plug is securely plugged into the power outlet to prevent re The power...

Page 5: ...harmful or radioactive particles If the refrigerator is to be decommissioned unplug the power cord to avoid electric shock current leakage or re caused by aged power lines If the refrigerator is left...

Page 6: ...accelerate the defrosting process other than those recommended by the manufacturer Do not damage the refrigerant circuit In order to reduce ammability HAZARDS the installation of this equipment shall...

Page 7: ...e as the temperature in the center of the refrigerator The cabinet temperature will gradually reach an equilibrium state Only clean the unit with light soapy water Never use brushes acids gasoline soa...

Page 8: ...lood bank refrigerator provided to you may be a little di erent from the gure in the manual due to product improvement The manual may be subject to changes without further notice 35 Name Blood Bank Re...

Page 9: ...frigerator Do not install the refrigerator in a narrow and con ned space as heat may be trapped to increase ambient temperature Additionally a unit should not be installed in a space where the entry d...

Page 10: ...14 1055 Interior dimensions W x D x H mm 460 370 950 430 350 830 E ective volume 158L 106L Door Glass door with electric heat Insulation CFC Free foamed in place urethane Compressor High Quality Herme...

Page 11: ...mal operation 6 Installation of the Handle Handle installation of products shall be conducted in the following steps Orient the handle with the hole downwards Align the handle and handle guide on door...

Page 12: ...ce the unit is powered up the audible alarm is turned on Turn on the battery switch and press the buzz cancel button to silence the alarm Visual alarm continues until the temperature in the measuremen...

Page 13: ...oducts Do not load in too much warm product at one time Overloading the refrigerator will reduce the system s expected life Therefore product and samples should be placed into the refrigerator in batc...

Page 14: ...g Temperature recorder Optional for HXC 158 Blood storage basket or drawer Shelf USB port optional Interior light Temperature measurement box Caster Door ajar alarm switch Air filter grille 29 Refrige...

Page 15: ...perature is too high Air duct is blocked Unit seems to emit too much noise Check the installation of the unit to make sure the unit does sit on a solid oor without vibration Whether certain part of th...

Page 16: ...temperature recorder is standard on the product optional for HXC 158 If additional chart paper is needed please contact Haier Biomedical The ordering cycle is 1 2 months Consult us by phone for detail...

Page 17: ...e after sales service personnel Cleaning and Maintenance Warning Clean the refrigerator Replace the illumination light Condensate evaporator is located in the back of the refrigerator Remove the 2 scr...

Page 18: ...tem is in a n alarm mode press the Buzzer Cancel key on the display board and the audible alarm is silenced temporarily Remote alarm condition will not be a ected After 30 minutes if the alarm conditi...

Page 19: ...timely manner to protect the samples When there is a fault in the battery E5 will show after pressing Alarm test Check the battery charge switch to make sure it is in the On position If it is O turn...

Page 20: ...ly Buzzing Alarm light ashes Display actual temperature Remote alarm contacts reverse and audible alarm starts Remote alarm can be cancelled after pressing Silence button If the alarm condition persis...

Page 21: ...y battery is low ashing of LED indicator lamp on the recorder indicates that the battery needs to be changed Maintenance and change of emergency battery of the recorder When the green LED of recorder...

Page 22: ...lay panel ashes USB which indicates the data are now exporting and not nished yet After 5 seconds the display will return to the temperature mode If the display shows ALL indicating the data export pr...